![Accu-Chek Inform 2 Operator'S Manual Download Page 52](http://html.mh-extra.com/html/accu-chek/inform-2/inform-2_operators-manual_455938052.webp)
Patient Glucose Testing
51
The following messages
1
may appear in addition to the
standard test result:
n
CR LO
(below the Critical Range threshold, but
within the Reportable Range)
n
CR HI
(above the Critical Range threshold, but
within the Reportable Range)
n
RR LO
(below the Reportable Range threshold, but
within the System Measurement Range)
n
RR HI
(above the Reportable Range threshold, but
within the System Measurement Range)
n
LO
(below the System Measurement Range)
n
HI
(above the System Measurement Range)
For test results that lie outside the critical or reportable
range, a message (up to 100 characters in length) can
be configured during setup. This message is then
displayed with the corresponding test results.
Instructions on how to add comments are provided in the
following section.
If comments are not set as required and you do not wish
to add a comment to the test result, touch
to return
to the
Main Menu
.
Test results are also saved when the meter is powered
off in the results screen or when it powers off
automatically.
Remove the test strip and dispose of it in accordance with
applicable regulations and directives for disposal of
potentially infectious samples and materials.
1
Results that are below or above the Critical Range threshold, but
within the Reportable Range can be configured to display as
either numeric results or as CR LO or CR HI. All other results
exceeding specified limit values will always be displayed as RR
LO, RR HI, LO, or HI.
Summary of Contents for Inform 2
Page 1: ...ACCU CHEK Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ......
Page 11: ... 10 This page intentionally left blank ...
Page 65: ...Glucose Control Testing 64 This page intentionally left blank ...
Page 91: ...Linearity Testing 90 This page intentionally left blank ...
Page 99: ...Proficiency Testing 98 This page intentionally left blank ...
Page 149: ...General Product Information 148 This page intentionally left blank ...
Page 167: ...Appendix 166 This page intentionally left blank ...
Page 191: ...Supplement for Observed Test Sequence 190 This page intentionally left blank ...
Page 197: ... 196 This page intentionally left blank ...
Page 198: ......