![Accu-Chek Inform 2 Operator'S Manual Download Page 51](http://html.mh-extra.com/html/accu-chek/inform-2/inform-2_operators-manual_455938051.webp)
Patient Glucose Testing
50
n
The System Measurement Range refers to the
measurement range of the system itself (test strips
and meter) and is the only range which cannot be
configured. With ACCU-CHEK Inform II test strips
this fixed range is 10 to 600 mg/dL. If a result falls
outside this range, the message HI or LO appears,
i.e., the result cannot be quantified properly.
n
Glucose results above or below the Reportable
Range, as defined by the institution or local
regulatory body, are above the highest or below the
lowest numerical results to be reported. Results
outside this range must not be used for
intervention decisions.
n
Glucose results above or below the Critical Range,
as defined by the institution, require immediate
action as defined by hospital policy.
n
Glucose results within the Normal Range, as
defined by the institution, are considered normal
and require no therapeutic action.
Summary of Contents for Inform 2
Page 1: ...ACCU CHEK Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ......
Page 11: ... 10 This page intentionally left blank ...
Page 65: ...Glucose Control Testing 64 This page intentionally left blank ...
Page 91: ...Linearity Testing 90 This page intentionally left blank ...
Page 99: ...Proficiency Testing 98 This page intentionally left blank ...
Page 149: ...General Product Information 148 This page intentionally left blank ...
Page 167: ...Appendix 166 This page intentionally left blank ...
Page 191: ...Supplement for Observed Test Sequence 190 This page intentionally left blank ...
Page 197: ... 196 This page intentionally left blank ...
Page 198: ......