background image

SIDE

   

Read instructions

before use.

OK-P

10-62-2965-0003  V1-FEB11

10-62-2965-0103      V1-FAB11

OKP010001

Before You Begin

PLEASE READ THIS BEFORE USING.

The following basic safety precautions should always be taken.

1.  Close supervision is necessary when the device is used by, on, or near children, 

handicapped persons or invalids.

2.  Use the device only for the intended use described in this manual.

3.  Do not use test strips and control solutions which are not supplied by the 

manufacturer.

4.  Do not use the device if it is not working properly, or if it has suffered any damage.

5.  Before using any product to test your blood glucose, read all instructions 

thoroughly and practice the test. Do all quality control checks as directed and 

consult with a diabetes healthcare professional.

KEEP THIS USER GUIDE WITH YOU.

Intended Use

The system is intended for use outside the body (in vitro diagnostic use only). It 

should be used only for self-testing blood glucose (blood sugar) and only with fresh 

capillary whole blood samples. The system is intended for use in the home and in 

clinical settings. The system should not be used for the diagnosis of diabetes or for 

the testing of newborns.

Principle of Measurement

Blood glucose is measured by an electrical current that is produced when a blood 

samples mixes with the reagent (special chemicals) of the test strip. The electrical 

     Getting To Know Your System

The OKmeter Cameo Blood Glucose Monitoring System.

The OKmeter Cameo system uses the latest technology to provide you with easy and 

comfortable testing. The system requires only a 0.7 μL of blood sample to complete 

the testing in only 6 seconds.

Important :

 Use only OKmeter Cameo test strips and control solutions with your 

OKmeter Cameo meter. Use other test strips and control solutions with this meter 

can produce inaccurate results. 

current changes with the amount of glucose in the blood sample. The OKmeter Cameo 

meter measures the strength of the electrical current, calculates your blood glucose 

level and then displays your result in either milligrams of glucose per deciliter (mg/dL) or 

millimoles of glucose per liter (mmol/L). 

Caution

1. The user should not take any decision of medical relevance without first consulting 

his or her medical practitioner. 

2.  Call your doctor immediately if you experience symptoms that are not consistent with 

your blood glucose test results.

3.  High altitudes above than 3,402 meter (11,161 ft) may affect the test results. 

4.  Temperatures outside the range of 10°C to 40°C (50°F to 104°F) may affect the test 

results. Do not test beyond of temperature range.

IMPORTANT HEALTH-RELATED INFORMATION

1.  Apply only capillary whole blood sample to test your blood glucose. Applying other 

substances or plasma, serum will cause wrong results.

2.  Severe dehydration and excessive water loss may cause false low results. If 

you believe you are suffering from severe dehydration, consult your healthcare 

professional immediately.

3.  Test  results  below  60  mg/dL  (3.3  mmol/L)*1  indicates  low  blood  glucose 

(hypoglycemia). Test results greater than 240 mg/dL (13.3 mmol/L)*2 indicates high 

blood glucose (hyperglycemia).If your results are below 60 mg/dL (3.3 mmol/L) or 

above 240 mg/dL (13.3 mmol/L), repeat the test, and if the results are still below 

60 mg/dL (3.3 mmol/L) or above 240 mg/dL (13.3 mmol/L), consult your healthcare 

professional immediately.

4.  Inaccurate results may occur in severely hypotensive individuals or patients in shock. 

Inaccurate low results may occur for individuals experiencing a hyperglycemic-

hyperosmolar state, with or without ketosis. Critically ill patients should not be 

tested with blood glucose meters.

5.  Abnormal red blood cell counts (hematocrit level below 20% or above 60%) may 

cause false results. Please consult your healthcare professional if you do not know 

your hematocrit level.

6.  Interference: Reducing substances occurring in the blood naturally (uric acid, 

bilirubin) or from therapeutic treatments (ascorbic acid, acetaminophen) will not 

significantly affect OKmeter Cameo test results. However, elevated concentrations 

of these substances may affect test results. The compounds listed in the table 

were found to have no affect at the concentration indicated.

Compounds

Highest concentrations 

tested at which no 

interference occured

Compounds

Highest concentrations tested 

at which no interference 

occured

Acetaminophen

 8.0 mg/dL (0.53 mmol/L)

Gentisic Acid

 5.0 mg/dL (0.32 mmol/L)

Ascorbic Acid 

 5.0 mg/dL (0.28 mmol/L)

Hydroxyurea

 3.0 mg/dL (0.39 mmol/L)

Aspirin

 60 mg/dL (3.33 mmol/L)

L-dopa

 10 mg/dL (0.51 mmol/L)

Bilirubin

 90 mg/dL (1.54 mmol/L)

Maltose

 900 mg/dL (26.3 mmol/L)

Cholesterol

 500 mg/dL (12.9 mmol/L)

Methyldopa

 3.0 mg/dL (0.13 mmol/L)

Creatinine

 5.0 mg/dL (0.44 mmol/L)

Tolbutamide

 400 mg/dL (14.8 mmol/L)

Dopamine

 2.0 mg/dL (0.11 mmol/L)

Triglycerides

 2,000 mg/dL (22.6 mmol/L)

Galactose

 900 mg/dL (50 mmol/L)

Uric Acid

 8.0 mg/dL (0.48 mmol/L)

REFERENCE:

* 1: Kahn, R. and Weir, G.: Joslinis Diabetes Mellitus, 13thed Philadelphia : Lea and Febiger (1994), 489.  
* 2: Krall, L.P. and Beaser, R. S.: Joslin Diabetes Manual. Philadelphia : Lea and Febiger(1989), 261-263.

1.

 

DATE: 

Display month and date.

2. TIME: 

The meter is programmed for a 12

    hour period format.

3. AVERAGE:

 

Appears when the meter is 

    in the memory mode while recalling 

    7/14/28 day test averages.

4. MEMORY SYMBOL: 

Appears when in

    the memory mode.

5. THERMOMETER SYMBOL: 

Appears 

    when ambient temperature is above or

    below the acceptable range needed for 

    testing. 

  6.

 

CONTROL SOLUTION TEST SYMBOL:

 

Appears when you are in Control Solution 

      Mode. Your test result will not be stored in meter memory.

  7.

  

UNIT OF MEASURE:

 

Appears with the test result either in mg/dL or mmol/L. 

  8.  BATTERY SYMBOL:

 

Appears when battery is weak.

  9. BLOOD DROP SYMBOL:

 

Flashes when sample should be applied.

10. STRIP SYMBOL

Appears when the meter is turned on and waiting for inserting a 

      test strip.

SETTING TIME AND DATE

Please install battery first and set correct time and date before you begin to test.

STEP 1. Enter Setting Mode

Start with the meter off. Open the battery cover. Press SET BUTTON 

located at the right side of battery compartment. The meter will turn 

on and enter the setting mode.

STEP 2. Set the Year

The year appears with the number flashing. Press and release the M 

BUTTON to obtain the desired year. Press SET BUTTON to confirm 

and move to next MONTH setting.

STEP 3. Set the Month

The month appears with the number flashing. Press and release 

the M BUTTON to obtain the desired month. Press SET BUTTON to 

confirm and move to next DATE setting.

STEP 4. Set the Date

The date appears with the number flashing. Press and release the M 

BUTTON to obtain the desired date. Press SET BUTTON to confirm 

and move to next HOUR setting.

STEP 5. Set the Hour

The hour appears with the number flashing. Press and release the M 

BUTTON to obtain the desired hour. Press SET BUTTON to confirm 

and move to next minutes setting.

STEP 6. Set the Minute

The minute appears with the number flashing. Press and release 

the M BUTTON to obtain the desired minute. Press SET BUTTON to 

confirm and the meter will enter to next unit setting.

STEP 7. Set the Unit of Measurement

The existing unit (for example, mg/dL) will appear and flashing. If you 

do not want to change the unit, Press SET BUTTON to skip this step. 

If you want to change the unit, press and hold the M BUTTON for 3 

seconds to change the unit to mmol/L. Press SET BUTTON to confirm 

and the meter will enter to next delete all memory setting.

STEP 8. Delete Memory

When the dEL symbol and the flashing memory symbol appear on 

the display, you can choose to clear the memory. If you do not want 

to clear the memory, press the SET BUTTON again to skip this step. 

If you want to clear ALL memory, press and hold M BUTTON for 3 

seconds. The “---“ image will appear on the LCD screen to indicate 

that all memory has been deleted. 

STEP 9. Complete Setting

After deleting memory, the meter will display “OFF” before shut down. 

The meter setting is now completed.

IMPORTANT :

 

1.The time, date and unit of measurement can ONLY be changed in the setting 

mode. Therefore, when you perform a blood glucose testing, it is not possible to 

change those parameters.

2.Your meter displays 7, 14 and 28-day averages which you can access from the 

meter memory. These averages are calculated from your latest result obtained 

during the last 7, 14 and 28 days. 

3.Your meter can display test result in milligram per deciliter (mg/dL) or millimoles 

per liter (mmol/L). Your meter has been preset at the factory with the standard 

unit measurement used where you live (mg/dL) . When you turn off the meter or 

replace the battery, the unit will not be changed.

4.Use of the wrong unit of measure may cause you to misinterpret your blood 

glucose level, and may lead to incorrect treatment. Please always consult with 

your healthcare professionals before you reset the unit of measure.

1.

A

1

. Getting To Know Your System

2. Prepare For Blood Sampling

3. Performing Blood Test
4. Memory Recall
5. Control Solution Testing
6. Caring For Your Meter And Test Strip
7. System Specifications
8. About Alternative Site Testing (AST)
9. Display Messages And Problem-Solving Guide

SIDE

 

 

A  :

   

SIDE

  

B  :

   

Meter

Test Strip

ABSORBENT 

CHANNEL:

Apply a drop of 

blood here.

CONTACT BARS:

Insert this end into strip slot.

TEST STRIP HANDLE :

Hold this part to insert the test strip 

into the meter slot.

CONFIRMATION WINDOW:

This is where we confirm if enough blood is 

applied for has been drawn into strip.

    Prepare For Blood Sampling

Adjustable Lancing Device 

Your lancing device and lancets are used for obtaining

blood samples from the puncture site.

Lancet

Protective cap

2

.

Important Lancing Device and Lancets Information

1. 

Lancet is for single use only.

2. Keep lancing device and lancets clean.

3. Use caution when removing the used lancet from the device and when disposing the 

used lancet.

IMPORTANT : 

The meter and lancing device are for single patient use.

                     Do NOT share them with anyone including other family members !

                     Do NOT use on multiple patients !

REPLACING THE BATTERY

The meter comes with ONE Lithium batteries (battery type CR2032) Before you start, 

please install battery with the plus(+) side up.
Battery life will vary depending on usage, so always keep a spare battery on hand. 
The battery should last about 1000 tests or 12 months when testing 3 times a day. 
When the battery symbol appears on the meter display, battery is getting low. You will 

still be able to test with low battery, but you should replace it as soon as possible.
When battery symbol appears together with E-b on the display, the meter will no longer 

give results and you must replace the battery immediately. Please always have one 

spare battery with you to ensure that you can replace the battery anytime.

How to replace the battery

 

1. Make sure the meter is turned off. Let the front of the meter 

rest in the palm of your hand. Press the buckle on the battery 

cover and lift up to open the cover. Remove the battery.

 

2. With the plus (+) side up, place the battery (battery type 

CR2032) in the compartment and slide it upwards until it is 

locked into place. You should hear a beep to indicate the 

battery installed correctly. If not, please replace the battery 

correctly.

 

3. Close battery cover.  

NOTE: 

1. Replacing the battery does not affect the test result stored in memory. However the 

time and date may need to re-set.

2. As with all small objects, the battery should be kept away from small children as a 

safety precaution. If the battery is swallowed, seek medical assistance immediately.

3. Batteries might leak chemicals if not used for a long time. Remove the batteries if 

you are not going to use the device for extended period(i.e., 3 months or more ).

4. 

Please discard the product or the batteries properly according to the regulations 

of your country.

For 

in vitro diagnostic use.

Temperature limitation / Store at.

 

Please consult instructions 

for use.

 

Use by / Expiry date.

 

Do not reuse.

 

Manufacturer.

 

Lot number.

EU representative.

 

Caution, consult 

accompanying document.

 

This product fulfils the requirements 

of Directive 98/79/EC 

in vitro 

diagnostic medical device.

Performance Characteristics:

Accuracy: 

 95% of the individual glucose results fall within ± 15 mg/dL(0.83 mmol/L) at glucose concentration < 75 mg/

dL(or 4.2 mmol/L) and within ± 20% at glucose concentration 

75 mg/dL(or 4.2 mmol/L), compared with glucose analyzer 

YSI 2300.

PRECISION: CVs (%) of intermediate precision and repeatability were less than 5%.

DISPLAY :

Shows test results 

and messages.

METER LABEL :

Each meter has its 

unique number on it. 

Do not alter or tear 

the label off.

SET BUTTON :

Located in the battery 

compartment; press button to 

enter meter setting. Refer to 

"

SETTING TIME AND DATE 

section for details.

BATTERY 

COMPARTMENT :

Holds ONE 3V Lithium battery 

(battery type CR2032). 

STRIP SLOT :

Insert the test strip here. 

The meter will turn on 

automatically.

M BUTTON :

Main button, press 

button to turn the 

meter on or perform 

other functions 

described in User 

Guide.

DATA PORT :

USB port, allows you to transfer the information stored 

in the meter to a computer to view, analyze and print. 

1 5 6

2

3

8

10

7

9

4

Blood Glucose Monitoring System

It's time 

to

travel l

igh

t

OK Biotech Co., Ltd. 

            

No. 91, Sec. 2, Gong Dao 5th Road, 

30070 Hsinchu City, TAIWAN

TEL: +886-3-5160258

FAX: +886-3-5160028

www.okbiotech.com

E-mail: [email protected]

MDSS GmbH
Schiffgraben 41,
30175 Hannover, 

GERMANY

V2-DEC14

Setting your Lancing Device

 

1. Screw off the cap of lancing device. Insert a 

lancet into the lancet holder and push down until 

it is fully seated.

2. Twist off the protective cap until it separates from 

the lancet.

3. Replace the lancing device cap and set the 

puncture depth to the desired number.

NOTE: THE DEPTH SELECTOR OFFERS 5 LEVELS OF SKIN
1-2 : FOR SOFT OR THIN SKIN
3 : FOR AVERAGE SKIN
4-5 : FOR THICK OR CALLOUSED SKIN

 

4. Pull back the cocking control until it makes a 

click, and then release. If it does not click, the 

device may have been cocked when the lancet 

was inserted.

 

Lancing Device Cap

Lancet Holder

Depth Selector

Release Button

Cocking Control

Reviews: