5 Operation | 5.4 Setting Up for Transmitted Light Techniques
ZEISS
Procedure
1. If required, push the
TL button
for transmitted
light illumination.
2. Adjust the image brightness using the
Inten-
sity/LM knob
on the microscope stand.
3. Insert the high-contrast sample into the sample holder of the stage.
4. Set position H (or B = brightfield), when using
condensers with a turret/modulator disk and
knurled ring.
5. Swivel the front lens into the beam path with
≥ 10x objectives, if condensers with a swiveling
front lens are used.
88
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001