7 Troubleshooting | 7.1 Resetting the Microscope to the Factory Settings
ZEISS
Symptom
Cause
Measure
No voltage in the power
socket.
Use another power socket.
Halogen lamp 12 V, 50 W
flickers, illumination in-
tensity is not stable.
Halogen lamp 12 V, 50 W is
reaching the end of its life
span.
Exchange the halogen lamp
12 V, 50 W, see
12 V, 50 W Halogen lamp of
the HAL 50 Halogen Illumina-
tor [
.
Power cable is not installed
properly or is damaged.
Install the power cable prop-
erly or exchange it.
The pins of the halogen lamp
12 V, 50 W are not properly
inserted in the socket.
Insert the pins of the halogen
lamp 12 V, 50 W correctly, see
Replacing the 12 V, 50 W
Halogen lamp of the HAL 50
Halogen Illuminator [
.
7.1 Resetting the Microscope to the Factory Settings
NOTICE
Please use this function with caution as it will reset all existing configurations.
The default factory settings are:
§
The Light Manager is enabled, but no light intensity values are saved.
§
The light intensity is set to the initial minimum value.
§
All stored configurations are cleared.
Prerequisite
ü
The microscope is operational.
Procedure
1. Press and hold down the
Intensity/LM knob
for 20 seconds.
à
While the knob is held down from 3 s to 20 s, the indicator light blinks in RED.
à
After 20 s the indicator light blinks green.
9
When the indicator stops blinking and remains GREEN, the reset to the default factory setting
is successful.
134