ZEISS
4 Installation | 4.5 Assembling Objectives
4.5 Assembling Objectives
1
2
3
4
5
6
1
Dust protection cap
2
Objective
3
Nosepiece
4
Adapter W0.8/M27
5
Specimen marker
6
Protective cap
Procedure
1. Use the focus drive to fully lower the mechani-
cal stage or lower the stage carrier by loosen
the clamping handle.
2. Remove the dust protection caps
1
from the appropriate openings in the nosepiece.
3. Remove objectives
2
from the case and screw them into the nosepiece
3
.
4. Carefully screw the objective into the opening. Start with the smallest magnification factor
(set up clockwise) in nosepiece position 1.
5. Make sure it engages properly in the nosepiece's thread.
6. Instead of an objective, the sample marker
5
with an adapter W0.8/M27
4
can be
screwed on in any desired nosepiece position.
7. Apply the protective cap
6
to prevent sample marker from drying out.
8. Always replace the dust protection caps on any empty positions on the nosepiece.
NOTICE
Dust-sensitive components
If unused nosepiece openings remain uncovered, particles may enter the microscope and may
damage its optics and mechanics permanently.
4
Always close unused nosepiece openings with cover caps!
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001
67