5 Operation | 5.5 Setting Up for Reflected Light Techniques
ZEISS
3. Slide the according DIC slider
1
into the slot
of the nosepiece.
1
2
4. Place the sample on the stage.
5. Focus the sample.
6. Turn the mechanical stage so the structure of interest is visible with maximum contrast.
7. Use the knurled screw
2
on the DIC slider to adjust the optimal contrast.
5.5.4 Setting Up for Reflected Light C-DIC Microscopy
For more information about the method, see chapter
Reflected Light DIC and C-DIC Microscopy
Prerequisite
ü
A reflected light light source is
.
ü
The microscope is operational.
ü
C-DIC reflector module is installed.
ü
A suitable objective for DIC is
, e.g. EC Epiplan-Neofluar, Epiplan with the addi-
tional label "DIC" or "Pol".
ü
C-DIC slider compatible with the objectives in use, is available.
ü
The illumination is adjusted for
reflected light brightfield microscopy [
Procedure
1. Swivel the C-DIC module on the reflector turret
into the beam path.
2. Swivel the DIC compatible objective into the
beam path.