10.7.3 Setting the Drive Length on the Stage Drive
1
2
1
Coaxial knurled knob for the Y adjust-
ment
2
Coaxial knurled knob for the X adjust-
ment
Procedure
1. Expand the length of the X
2
and Y drives
1
by shifting the respective coaxial
knurled knob axially within a range of approx. 15 mm.
10.7.4 Removing the Additional Sleeves on the Ergonomic Stage Drive
Both coaxial knurled knobs for the stages are equipped with additional sleeves for an even more
sensitive adjustment of the sample position. These sleeves can be removed when a faster sample
movement is important.
Procedure
1. Loosen both clamping screws
6
on the
lower additional sleeve
5
.
1
2
3
4
5
6
2. Remove the lower additional sleeve
5
from the X-coaxial knurled knob
2
by pulling
it down.
3. Loosen both clamping screws
4
on the upper additional sleeve
3
.
4. Remove the upper additional sleeve
3
from the Y-coaxial knurled knob
1
by pulling
it down.
Proceed in the reverse order for installing.