10 Accessories and System Expansions | 10.10 Assembling the Sample Space Extension, 60 mm
ZEISS
10.10.3 Removing the Upper Stand Part
4
9
Fig. 82: Removing the upper stand part
4
Upper stand part
9
Fixing screw (6x)
Procedure
1. Unscrew 6 fixing screws
9
, holding the upper stand part
4
.
2. Carefully remove the upper stand part while pulling the cables downwards out of the upper
stand part.
10.10.4 Assembling the Sample Space Extension
10
11
7
12
5
Fig. 83: Installing the sample space extension
5
Cable connector of the main control
board
7
Cable connector of the main control
board
10
Spacer bolt (6x)
11
Sample space extension
12
Upper stand part
Procedure
1. Carefully route the cables
5
and
7
from below through the long hole of the sample
space extension
11
.
2. Place the sample space extension on the lower stand part
12
.
3. Tighten the sample space extension with its 6 spacer bolts
10
.
172
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001