ZEISS
8 Decommissioning and Disposal | 8.1 Decommissioning
8 Decommissioning and Disposal
This chapter contains information on the decommissioning and disposal of the microscope and its
expansions/components or accessories.
8.1 Decommissioning
If the microscope and its components are not used for an extended period such as several
months, they should be shut down completely and secured against unauthorized access.
DANGER
Electric shock due to live parts
When the microscope is still switched on, coming in contact with live parts can lead to electric
shock or burn.
4
Switch off the microscope prior to opening or cleaning.
4
Disconnect live parts from the power supply.
Procedure
1. Switch off the microscope.
2. Pull the mains plug.
3. Protect microscope using a dust cover.
8.2 Transport and Storage
The following regulations must be observed before and during transport:
§
The boxes must be secured during transport.
§
Avoid rocking the boxes back and forth.
§
Note the weight information on the package and on the shipping document.
§
Where possible, the original packaging must be used for shipping or transport.
Maximum shock
resistance
§
Do not drop or bump the boxes during movement or storage. Any acceleration shall be
< 10 g.
§
Evaluate packaging shock and tilting sensors on delivery and after internal transport.
Allowable
temperature
Allowable temperature during transport in packaging:
§
Between -40 °C and +70 °C
§
Relative humidity (without condensation) less than 75 % at 35 °C
Allowable temperature during storage:
§
B10 °C and +40 °C
§
Relative humidity (without condensation) less than 75 % at 35 °C
Info
24 hours before installation
of the microscope it is required that the boxes be at recom-
mended room temperature to avoid ingress of humidity, which is very harmful to optical paths,
and to ensure effective stability of the microscope during installation and testing.
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001
135