10 Accessories and System Expansions | 10.15 Assembling the Base Plate on the Stand
ZEISS
5. If the overlap is not exact
A
, recenter the
lighter annular diaphragm.
6. Remove the auxiliary microscope and replace the eyepiece.
10.15 Assembling the Base Plate on the Stand
2
1
3
4
Fig. 106: Installing the base plate on the stand
1
Stand
2
Rubber foot (4x)
3
Base plate
4
hexagonal socket screw (4x)
Procedure
1. Remove the four rubber feet
2
from the bottom of the stand
1
.
2. Align the base plate
3
with the stand.
3. Insert four hexagonal socket screws (M6)
4
into the holes in the base plate.
4. Tighten the screws.
Proceed in the reverse order for removal.