![Zeiss Axioscope 5 Скачать руководство пользователя страница 78](http://html2.mh-extra.com/html/zeiss/axioscope-5/axioscope-5_instruction-manual_3365198078.webp)
4 Installation | 4.12 Assembling the Reflected Light Light Source
ZEISS
4.12.4 Assembling the HBO 100 Light Source
CAUTION
Hazardous light emission
The light source belongs to Risk Group 2 as specified in IEC 62471 and emits LED radiation.
Eye damage may result from exposure.
4
Never look directly into the light-emitting aperture of the light source.
4
Before installing or removing the light source always make sure it is switched off.
NOTICE
Risk of damage to the gray filter
The high light intensity of the light source can damage the gray filter for reflected light during
prolonged use.
4
Use an attenuator instead of a gray filter to change the light intensity in the reflected light
path.
Info
The light source cannot be mounted onto the standard stand.
When using the HAL or HBO light sources, a base plate for the Axioscope (000000-2202-526)
is mandatory.
Info
For installing or replacing the HBO 103 W/2 mercury vapor short-arc bulb at the HBO 100 light
source, consult the operator manual supplied with the light source.
Parts and Tools
Hex Key, 3.0 mm
Prerequisite
ü
The microscope is switched off.
ü
The base plate is installed. [
ü
The HBO 100 light source is switched off.
ü
The HBO 103 W/2 mercury vapor short-arc bulb is installed at the light source.
ü
The stand is equipped with the illumination mount.
ü
The protective cap is removed from the illumination mount of the stand.
Procedure
1. Loosen the clamping screw
2
on the illumi-
nation mount
3
.
3
2
1
2. Insert the light source with the dovetail ring into the illumination mount.
3. Fasten the clamping screw.
78