4 Installation | 4.11 Assembling the Transmitted Light Light Source
ZEISS
4.11 Assembling the Transmitted Light Light Source
4.11.1 Assembling the LED10 Light Source for Transmitted Light Illumination
CAUTION
Hazardous light emission
The light source belongs to Risk Group 2 as specified in IEC 62471 and emits LED radiation and
UV radiation. Eye damage or skin irritation may result from exposure.
4
Never look directly into the light-emitting aperture of the light source.
4
Avoid exposure of skin to the radiation. Use suitable protective equipment/protective
clothing if required.
4
Before installing or removing the light source always make sure it is switched off.
1
2
3
5
4
6
2
Fig. 45: Assembling the LED10 light source for transmitted light illumination
1
3 mm hex key
2
Clamping screw
3
LED10 light source
4
Plug
5
Socket for transmitted light illumination
6
Illumination mount
Parts and Tools
Hex Key, 3.0 mm
Prerequisite
ü
The microscope is switched off.
ü
The stand is equipped with the illumination mount.
Procedure
1. At the illumination mount on the rear side of the stand, loosen the clamping screw
2
.
2. Insert the plug
4
of the light source's cable into the socket
5
for transmitted light il-
lumination.
3. Insert the light source's dovetail ring into the illumination mount
6
.
4. Ensure that the illumination cable is not being pinched or clamped.
5. Tighten the clamping screw
2
.
Proceed in the reverse order for removal.
74