6 Care and Maintenance | 6.3 Maintenance Work
ZEISS
Prerequisite
ü
The microscope is switched off.
ü
The plug of the light source's cable has been removed from the corresponding socket.
ü
The light source has cooled down for about 15 minutes.
Procedure
1. Press the release button
7
and pull the bulb carrier
6
completely out to the side.
2. Put the bulb replacement tool
2
onto the old bulb
3
.
3. Press down the two spring levers
4
and remove the tool with the bulb in upward direc-
tion.
4.
NOTICE
Do not touch the new bulb with bare hands!
Put the bulb replacement tool onto the replacement bulb
5
.
5. Press down the two spring levers and insert the new bulb into the bulb socket
8
.
6. To center the bulb, briefly press the spring levers once more.
7.
NOTICE
Do not leave the bulb replacement tool inside the light source.
Insert the lamp carrier into the Hal 100 housing
1
and slide it in until it clicks into place.
6.3.4.2 Adjusting the HAL 100 Light Source
CAUTION
Burning hazard due to hot light sources
Light sources can become hot during processing.
4
Avoid touching the hot light source housing.
4
Let the light source cool down before touching it.
CAUTION
Eye injury due to light emission
Directly looking into emitted light can damage the eye.
4
Do not look into the light exit aperture of the light source.
The following action comprises several action sequences. These sequences are to be carried out in
the specified order.
§
§
122
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001