5 Operation | 5.4 Setting Up for Transmitted Light Techniques
5.3.6.3 Disabling the Light Manager Function
Prerequisite
ü
The microscope is operational.
ü
.
Procedure
1. Press one of the
Snap button
and the
Inten-
sity/LM knob
simultaneously for at least 1.5
seconds.
à
The indicator light blinks in the following sequence: GREEN / ORANGE / GREEN
5.3.7 Setting the ECO/Permanent Mode
Prerequisite
ü
The microscope is operational.
Procedure
1. Select the ECO or Permanent mode for micro-
scope illumination using the
ECO/Permanent
mode switch
.
5.4 Setting Up for Transmitted Light Techniques
5.4.1 Setting Up for Transmitted Light Brightfield Microscopy Using the KÖHLER Method
§
High-contrast sample slide
Every microscope (except the one with the Vario stand column) is configured to work with the
transmitted light brightfield method. All available condensers (except special condensers like dark-
field condensers) can be used for the transmitted light brightfield method.
Prerequisite
ü
The microscope is operational.
ü
The height stop of the condenser carrier is
ü
The height stop of the focusing drive is
ü
A suitable condenser for TL brightfield microscopy is installed.
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001