ZEISS
2 Safety | 2.1 Intended Use
2.1.2 Lifetime
A microscope is an opto-electronic device. Its availability for use is significantly determined by the
performed maintenance. ZEISS guarantees the ability for maintenance and repair within eight
years after initial operation. This is ensured by a corresponding service and spare parts concept,
thus enabling the intended purpose within this duration.
2.1.3 Optical Risk Grouping
According to EN 62471 sources of optical radiation are classified into risk groups subject to their
potential photobiological hazard. Sources are classified into the following four groups according
to hazard, based on the emission limit as well as permissible exposure time before hazard ex-
ceeded.
Risk group
Description
Exempt
No photobiological hazard.
1
No hazard due to normal behaviourial limitations on exposure.
2
No hazard due to the aversion response to very bright light sources or
thermal discomfort.
3
Hazardous even for momentary exposure.
The following table lists the risk grouping of the available light sources/illumination units accord-
ing to the mentioned standard:
Light source/Illumination unit
Risk group
Colibri 3
3 (high risk)
HBO 100
2 (moderate risk)
HXP 120
2 (moderate risk)
LED modules
2 (moderate risk)
HAL 100
2 (moderate risk)
HAL 50
1 (low risk)
See also
2
Usability of LED Modules for the Colibri 3 LED Light Source [
2.1.4 EMC Information
The microscope is intended to be used in a basic electromagnetic environment. The electromag-
netic environment should be evaluated prior to operation of the microscope. Do not use the mi-
croscope in proximity to sources of strong electromagnetic radiation, as these can interfere with
proper operation.
Use of this microscope in a dry environment, especially if synthetic materials are present (synthetic
clothing, carpets, etc.), may cause electrostatic discharges that may cause erroneous results.
Electromagnetic interference (EMI) according to CISPR 11 Group 1:
§
Class A (only Axioscope 5 Vario)
§
Class B (all other Axioscope 5 and Axioscope 7 stands)
If in doubt, contact a ZEISS service representative.
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001
15