ZEISS
6 Care and Maintenance | 6.3 Maintenance Work
Parts and Tools
Hex Key, 3.0 mm
Prerequisite
ü
The microsope is switched off.
ü
The plug of the light source's cable has been removed from the corresponding socket.
ü
The Colibri 3 light source is removed from the microscope.
Procedure
1. Loosen the four captive screws
5
on the front side
6
of the light source.
2. Remove the housing
7
.
3. Disconnect the LED module power cable connector
1.1
from the PCBA
1.6
.
4. Loosen the knurled screw
1.2
.
5. Remove the old LED module
1.4
.
6. Select the LED module with matching LED-specific labels
1.3
and
1.5
.
7. Insert the LED module in the correct slot.
8. Connect the LED module power cable connector to the PCBA.
9. If required, replace the LED modules of LED slots 2, 3 and 4 in the same way.
10. Re-mount the housing.
6.3.6.1 Usability of LED Modules for the Colibri 3 LED Light Source
Position
Slot 1
Slot 2
Slot 3
Slot 4
Wavelength range [nm]
450-480
350-415
594-660
508-565
385 nm LED (423052-9593-000)
X
O
X
X
470 nm LED (423052-9573-000)
O
X
X
X
505 nm LED (423052-9562-000)
X
X
X
O
565 nm LED (423052-9602-000)
X
X
X
O
625 nm LED (423052-9522-000)
X
X
O
X
O = usable
X = not usable
6.3.7 Recovering the Stage Travel Range in X-Axis
After long hours of use, the X direction range will gradually become smaller. This is not a quality
issue and can be easily reset.
Procedure
1. Hold the sample holder's two screws
1
/
2
.
1
2
2. Move the stage to the left until it hits the end stop.
3. Move the stage to the right until it hits the end stop.
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001
127