ZEISS
10 Accessories and System Expansions | 10.7 Stages
10.7.2.1 Assembling the Pol Rotary Stage
1
2
3
6
4
5
Fig. 74: Installing the Pol rotary tage
1
Pol rotary stage
2
Clamping device (2 stage clips)
3
Knurled screw for locking the rotation
4
Spring-loaded pin
5
Screw cap
6
Stage carrier for rotary stages
Procedure
1. Loosen the screw cap
5
of the spring box with approx. 3 turns.
2. Put the rotary
1
stage with the dovetail ring notch onto the spring-loaded pin
4
.
The knurled screw
3
must point to the front right.
3. Press the stage forward against the spring-loaded pin and lower it towards the back into
the stage carrier
6
.
4. Tighten the screw cap.
5. Insert the stage clips of the clamping device
2
into the holes on the stage provided for
this purpose.
Proceed in the reverse order for removal.
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001
159