10 Accessories and System Expansions | 10.11 Polarizers
ZEISS
10.11.6.1 Assembling Polarizer or Color Filter Carrier on the Condenser Carrier
The following polarizers or the filter carrier can be installed on the condenser carrier:
§
Polarizer D, fixed, removable
§
Polarizer D, 90° rotary, removable
§
Polarizer, rotary, with color filter carrier
§
Polarizer, fixed, with lambda plate, rotary
§
Circular polarizer D, fixed, with rotary lambda/4 plate
§
Circular polarizing equipment D ACR, with rotary lambda/4 plate
§
Color filter carrier 3x for filter d=32 mm
1
2
3
4
3
5
1
Condenser carrier
2
Locking pin
3
Adjusting lever
4
Holding pin
5
Polarizer
Prerequisite
ü
The stage carrier with the condenser carrier is
ü
Procedure
1. Hold the polarizer
5
parallel to the bottom of the condenser carrier
1
.
2. Insert the holding pin
4
into the front threaded opening at the left below the condenser
carrier.
3. Tighten the holding pin with the adjusting lever
3
.
4. Screw the locking pin
2
with the adjusting lever into the rear threaded opening of the
condenser carrier.
Proceed in the reverse order for removal.
10.11.7 Low-power System for Objectives 2.5x/4x
Purpose
The low-power system is for full display field illumination when using an objective with a weak
magnification factor (2.5x–4x) in combination with the Condenser 0.9/1.25 H.
Position
The low-power system is mounted behind the condenser carrier.
Function
It can be centered and remains swivelled into the beam path for as long as the respective objec-
tive is in use.
The illumination of weak objective magnifications can be centered with the centering screws. For
this purpose, the condenser should be centered on the other objectives without the low-power
system.
178