5 Operation | 5.3 Adjusting
ZEISS
5.3.6.2 Saving Light Intensity Ratios Using the Light Manager Function
Prerequisite
ü
The microscope is operational.
ü
The Light Manager function is
Procedure
1. Switch to the first objective and/or reflector (if
available) positions of interest using the knurled
rings (or slider).
2. Set the desired light intensity by turning the
In-
tensity/LM knob
.
3. Press the
Intensity/LM knob
for at least 1.5 seconds.
à
The light intensity for this objective/reflector combination is saved.
à
When using LED as light source, the LED is switched off for 300 ms. This is visible
through the eyepieces and serves as an indicator for the user.
4. Switch to the second objective/reflector position.
5. Press the
Intensity/LM knob
knob for at least 1.5 seconds.
à
Now a ratio between the first and the second objective/reflector combinations is estab-
lished.
6. Repeat to set light intensity ratios for more objective/reflector combinations.
9
After switching on the microscope, the previous setting of the Light Manager will be restored.
See also
2
Enabling the Light Manager Function [
86
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001