10 Accessories and System Expansions | 10.11 Polarizers
ZEISS
10.11.7.1 Assembling and Centering the Low-Power System
Info
The low-power system can only be used in combination with the condenser 0.9/1.25.
3
6
5
1
3
2
4
4
1
Condenser carrier
2
Locking pin
3
Adjusting lever
4
Adjusting screw (2x)
5
Holding pin
6
Low-power system
7
Handle of the low-power system for
swivelling in/out
Parts and Tools
2 x Hex key, 1.5 mm
Prerequisite
ü
The stage carrier with the condenser carrier is
ü
The condenser or filter holder is
Procedure
1. Hold the low-power system
6
parallel to the underside of the condenser carrier
1
and screw the holding pin
5
of the low-power system with the angled adjusting lever
3
into the front left threaded hole below the condenser carrier as far as it will go.
2. Screw the locking pin
2
with the adjusting lever as far as it will go into the rear
threaded hole of the condenser carrier.
3. Swivel the low-power system into the beam path using the handle
7
.
à
Make sure that it is securely engaged.
4. Switch on the microscope.
5. Set transmitted light illumination.
6. Open the aperture diaphragm and luminous-field diaphragm completely.
7. Adjust both adjustment screws
4
until the field of vision is optimally illuminated.
Proceed in the reverse order for removal.
180