10 Accessories and System Expansions | 10.10 Assembling the Sample Space Extension, 60 mm
ZEISS
10.9.3 Changing PhC DIC PlasDIC Diaphragms on the 0.9 BF DF PhC DIC Achromatic-Aplanatic
Condenser
1
1
2
Fig. 79: Changing PhC DIC PlasDIC diaphragms
1
Knurled ring of the modulator disk
2
Cover
Parts and Tools
Hex key, 1.5 mm
Prerequisite
ü
The condenser is
Procedure
1. Remove the cover
2
.
2. Turn the modulator disk
1
to have access to the corresponding diaphragm position.
3. Change the appropriate diaphragm, see
Assembling the Slit-Diaphragm for PlasDIC into the
4. If a DIC diaphragm has been inserted instead of a PhC diaphragm, the pre-set mechanism
that automatically opens the PhC diaphragm must be deactivated. For this, turn the center-
ing screws of the modulator disk counterclockwise to the stop.
à
Now, the aperture diaphragm can be closed for DIC contrast techniques.
10.10 Assembling the Sample Space Extension, 60 mm
Info
Pure transmitted light stands have no reflector turret sensor / RL LED interface. Accordingly, no
connection cables for this purpose are installed.
The following action comprises several action sequences. These sequences are to be carried out in
the specified order.
1.
Removing the cover of the upper stand part [
2.
Disconnecting the cable connections [
3.
Removing the upper stand part [
4.
Installing the sample space extension [
5.
Installing the upper stand part on the sample spaceextension [
6.
Establishing the cable connections [
7.
Installing the cover of the upper stand part [
170