ZEISS
10 Accessories and System Expansions | 10.9 Loading the Condenser
10.9.2 Assembling the Slit-Diaphragm for PlasDIC into the Modulator Disk
1
2
3
4
5
6
7
8
Fig. 78: Installing the slit-diaphragm for PlasDIC
1
Tool
2
Adapter ring for slit-diaphragm
3
Slit-diaphragm for PlasDIC
4
Adapter ring for PhC diaphragm
5
Spring
6
PhC diaphragm
7
Centering screws
8
Orientation grooves
Parts and Tools
Hex key, 1.5 mm
Prerequisite
ü
.
Procedure
1. Turn the PhC diaphragm that needs to be replaced into the free aperture of the modulator
disk.
2. Screw the modulator disk’s centering screws
7
back until they stop.
3. Unscrew the adapter ring
4
of the PhC diaphragm with the included tool
1
.
4. Remove the PhC diaphragm
6
and the spring
5
.
5. With the tool, insert the slit-diaphragm for PlasDIC
3
with its cams into the orientation
grooves
8
.
6. With the tool, screw in the adapter ring
2
included with the slit-diaphragm.
Proceed in the reverse order for removal.
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001
169