10 Accessories and System Expansions | 10.10 Assembling the Sample Space Extension, 60 mm
ZEISS
10.10.6 Establishing the Cable Connections
5
6
7
8
Fig. 85: Establishing the cable connections
5
Cable connector of the main control
board
6
Cable connector of the turret sensor /
RL LED interface
7
Cable connector of the main control
board
8
Cable connector of the nosepiece sen-
sor
Procedure
1. Establish the cable connection between the main control board
7
of the lower stand
part and the nosepiece sensor of the upper stand part
8
.
2. Establish the cable connection between the main control board
5
and the turret sen-
sor / RL LED interface
6
.
3. Make sure the plugs are securely connected.
10.10.7 Assembling the Cover of the Upper Stand Part
1
2
3
4
Fig. 86: Installing the cover
1
Screw (4x)
2
Cover
3
Cable tie
4
Upper stand part
Procedure
1. Secure the cables with a cable tie
3
.
2. Place the cover
2
on the upper stand part
4
.
3. Screw in and tighten the 4 screws
1
.
4.
174
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001