4 Installation | 4.10 Assembling the Dry Darkfield Condenser
4.10 Assembling the Dry Darkfield Condenser
5
7
8
4
1
2
3
6
Fig. 44: Assembling the darkfield condenser
1
Fastening ring
2
Darkfield condenser
3
Condenser holder Z
4
Mainspring
5
Centering screw (left/right)
6
Stage carrier
7
Condenser carrier
8
Knurled knob for vertical adjustment of
the condenser carrier
Procedure
1. Carefully move the stage carrier
6
with the
focusing drive to the upper stop position.
NOTICE
Damage due to collision. Make
sure that the stage does not collide with
the objective.
2. Using the knurled knob
8
for vertical adjustment, push the condenser carrier down as
far as it will go.
NOTICE
Damage due to collision. If using a low-power system, make sure that
this does not come to rest on the luminous-field diaphragm.
3. Unscrew both centering screws
5
on the condenser carrier
7
until their ends are no
longer visible.
4. Insert the darkfield condenser
2
in the condenser holder Z
3
.
5. Fix the darkfield condenser with the fastening ring
1
.
6. Press the condenser holder Z with the dovetail ring against the mainspring
4
of the con-
denser carrier until the condenser holder Z sits horizontally on the condenser carrier.
7. Screw in the centering screws until they engage with the dovetail ring of the condenser
holder Z.
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001