5 Operation | 5.3 Adjusting
5.3.2 Adjusting for Ametropia when Using Eyepiece Reticles
Prerequisite
ü
Two adjustable eyepieces are installed
ü
One eyepiece with reticle is installed.
Procedure
1. Focus on the line figure of the eyepiece reticle with the focusable eye lens of the adjustable
eyepiece containing the eyepiece reticle.
2. Focus on the microscopic image of a loaded sample with the focusing drive while observing
with the eyepiece containing the eyepiece reticle.
à
Both the microscopic image and the eyepiece reticle are in focus now.
3. Focus the microscopic image for the second eye with the focusable eye lens of the second
eyepiece.
9
Both microscopic images including the eyepiece reticle are thus in focus.
From this point, use only the focusing drive for any subsequent focusing activity.
5.3.3 Adjusting the Upper Stand Part Vertically
The present section applies to the following microscope type:
§
Axioscope 5 Vario (430035-9150-000)
The height of the upper stand part can be adjusted according to the size of the sample.
1
2
3
Fig. 49: Adjusting the upper stand part vertically
1
Release lever
2
Handwheel to adjust the height
3
Stand column
Procedure
1. On the stand column
3
, loosen the release lever
1
.
2. Use the handwheel
2
to adjust the height.
3. Tighten the release lever.
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001