ZEISS
4 Installation | 4.11 Assembling the Transmitted Light Light Source
4.11.2 Assembling the HAL 100 Light Source for Transmitted Light Illumination
CAUTION
Hazardous light emission
The light source belongs to Risk Group 2 as specified in IEC 62471 and emits LED radiation and
UV radiation. Eye damage or skin irritation may result from exposure.
4
Never look directly into the light-emitting aperture of the light source.
4
Avoid exposure of skin to the radiation. Use suitable protective equipment/protective
clothing if required.
4
Before installing or removing the light source always make sure it is switched off.
NOTICE
Heat damage
The HAL 100 bulb replacement tool might suffer damage from the emitted heat during the
light source operation.
4
Remove the bulb replacement tool from the HAL 100 housing before installing the light
source.
4
Do not operate the light source with the bulb replacement tool inside its housing.
Info
The light source cannot be mounted onto the standard stand.
When using the HAL or HBO light sources, a base plate for the Axioscope (000000-2202-526)
is mandatory.
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001
75