6 Care and Maintenance | 6.3 Maintenance Work
6.3.4 HAL100 Light Source
6.3.4.1 Replacing the Halogen Bulb 12 V, 100 W
CAUTION
Hazardous light emission
The light source belongs to Risk Group 2 as specified in IEC 62471 and emits LED radiation and
UV radiation. Eye damage or skin irritation may result from exposure.
4
Never look directly into the light-emitting aperture of the light source.
4
Avoid exposure of skin to the radiation. Use suitable protective equipment/protective
clothing if required.
4
Before installing or removing the light source always make sure it is switched off.
CAUTION
Burning hazard due to hot light sources
Light sources can become hot during processing.
4
Avoid touching the hot light source housing.
4
Let the light source cool down before touching it.
The light source does not have to be removed from the microscope for replacing the bulb.
1
2
3
5
6
7
8
4
6
Fig. 53: Replacing the HAL 100 light source bulb
1
HAL 100 housing
2
Bulb replacement tool
3
Old bulb
4
Spring lever (2x)
5
Replacement bulb
6
Bulb carrier
7
Release button
8
Bulb socket