INSTRUCTIONS

BX63

INTELLIGENT MICROSCOPE

This instruction manual is for the Olympus Intelligent Microscope Model BX63.

To ensure the safety, obtain optimum performance and to familiarize yourself fully with the use 

of this microscope, we recommend that you study this manual thoroughly before operating the 

microscope.

Retain this instruction manual in an easily accessible place near the work desk for future reference. 

A X 7 9 3 3

Содержание BX63

Страница 1: ...X63 To ensure the safety obtain optimum performance and to familiarize yourself fully with the use of this microscope we recommend that you study this manual thoroughly before operating the microscope...

Страница 2: ...in which case the user will be required to correct the interference at his own expense FCC WARNING Changes or modifications not expressly approved by the party responsible for compliance could void t...

Страница 3: ...4 19 BX63 4 8 Objectives with Correction Collar 32 2 Moving the Specimen 2 Adjusting the Diopter 1 Placing the Specimen 1 Adjusting the Interpupillary Distance 1 Centering the Condenser 2 Compatibilit...

Страница 4: ...8 OPTICAL CHARACTERISTICS UIS2 Series 38 9 ASSEMBLY See this section for the replacement of the halogen light bulb 39 44 11 SET TYPE LIST 46 48 n PROPER SELECTION OF THE POWER SUPPLY CORD 49 50 2 Sele...

Страница 5: ...Never insert metallic objects into the air vents of the microscope frame as this could result in electrical shock personal injury and equipment damage 6 Always use the power cord provided by Olympus I...

Страница 6: ...shock 3 The standard service life of the lamp housing is eight 8 years of use or 20 000 hours of total power ON period whichever is the shorter period For details see Inspection Sheet on page 45 4 To...

Страница 7: ...instruction manual Indicates that care is required against your finger or hand being caught Indicates that the main switch is ON Indicates that the main switch is OFF Caution indications Caution labe...

Страница 8: ...arding to a remote location contact Olympus to obtain advice for guaranteeing the performance CAUTION 2 Maintenance and Storage 1 To clean the lenses and other glass components simply blow dirt away u...

Страница 9: ...evice in the following country US Safety and EMC Precautions This product complies with the emission and immunity requirements described in IEC61326 series The electromagnetic environment should be ev...

Страница 10: ...es marked also read their instruction manuals Observation tube U BI30 2 U TBI3 CLI U TR30 2 U TTR 2 Eyepieces WHN10X WHN10X H Motorized 7 position nosepiece Coded 7 position nosepiece U D7REA U D7RES...

Страница 11: ...itted light observation combination Interpupillary distance adjustment scale Page 27 Diopter adjustment ring Page 27 Slide holder Page 22 Dummy cover Filter mount Page 20 Display panel See next page f...

Страница 12: ...te instruction manual for BX3 CBH U MCZ The manual shows the wiring diagrams for the connection cables DIP switch For use in operation setting Main switch I ON OFF Indicator LEDs POWER Power status in...

Страница 13: ...te instruction manuals for the BX3 RFAA and U AW Motorized Fluorescence Illuminator BX3 RFAA Lamp Housing for 100W Mercury Burner U LH100HG U LH100HGAPO Lamp Housing for 75W Xenon Burner U LH75XEAPO M...

Страница 14: ...instruction manual for BX3 CBH U MCZ Coarse adjustment knob Fine adjustment knob Fluorescence brightness control buttons st SNAP still image capture button LIVE PAUSE button Observation mode selector...

Страница 15: ...stage on the BX63 position it so that the stage knobs are on the right Note that the BX3 UCD8A motorized universal condenser cannot be used when this stage is installed CAUTION Stage center plate Y ax...

Страница 16: ...age button P 19 9 Objective button 10X P 19 a Brightness tab P 19 b Focus buttons sliders P 19 cBinocular tube P 27 dDiopter adjustment ring P 27 eCondenser height adjustment knob P 30 fCondenser cent...

Страница 17: ...ed fluorescence illuminator BX3 RFAA Motorized attenuator wheel U AW Motorized 7 position nosepiece U D7REA Control box for BX3 SSU Control box BX3 CBH Touch panel controller Motorized universal conde...

Страница 18: ...ed by a finger wearing a rubber or vinyl glove So use a touch pen for a static capacitance type touch panel 2 The selected button turns blue to indicate that it is selected 3 For the differences in th...

Страница 19: ...ch panel controller are already ON and the touch panel controller is in the standby mode the display starts with the blue display 5 Touch the Help button with your fingertip First read the following d...

Страница 20: ...Fig 7 Unit Setting This tab is used to set the units connected to the BX63 1 Tap the Unit tab 1 in the System Setting display to open the Unit display Fig 8 2 Set the units connected to the BX63 When...

Страница 21: ...tions 6 If you want to set another item tap the tab for the desired setting If you want to exit the system setting tap the X marking to close the System Setting screen Fig 9 Fig 11 3 3 4 4 2 2 1 1 Mir...

Страница 22: ...tting is required for the unused turret position numbers 6 If you want to set another item tap the tab for the desired setting If you want to exit the system setting tap the X marking to close the Sys...

Страница 23: ...1 Tap the desired mode in the MENU display Guidance The basic settings of microscope are performed automa tically with simple operations Full Operation In addition to the functions available with the...

Страница 24: ...the filter mount When using a halogen bulb be always sure to use the 45 LBD IF color temperature conversion filter CAUTION Fig 18 Using the Filter Cassette Figs 19 to 22 Loading Filters into Filter C...

Страница 25: ...ighten the screw to fasten the filter cassette When the filter cassette is installed the condenser may hit the filter cassette when lowered Therefore exercise caution when lowering the condenser with...

Страница 26: ...lamping lever Microscopy with Single Slide Holder Fig 24 The specimen slide can easily be placed by sliding it into the slide holder from the front Examining a whole specimen Use one of the thin slide...

Страница 27: ...ecimen may sometimes be unable to be focused with the condenser To prevent this remove the stopper screw on the mounting section of the BX3 STAD stage adapter in advance Fig 26 In this case certain co...

Страница 28: ...holder and move the stage guide to the left and right so that it hits the stoppers Y axis direction Hold the upper stage and move it to the front and rear so that it hits the stoppers Stage Feed Knob...

Страница 29: ...n carrying the microscope in order to prevent damaging it Fig 30 Fig 31 2 Moving the Specimen Fig 30 1 The specimen can be moved by turning the X axis knob and Y axis knob 2 The movement strokes are 5...

Страница 30: ...ig 33 5 Using the Light Shielding Sheet Fig 33 The light shielding sheet provided with the reflected light flu orescence illuminator is too small to be used with the BX63 Always use the light shieldin...

Страница 31: ...n slider of the touch panel controller Do not use an oil immersion objective 3 Engage a low power objective 10X or so in the light path rotate only the right diopter adjustment ring to bring the sampl...

Страница 32: ...position Slide the light path selector knob to select the desired light path Trinocular tube Light path selector knob position Pushed in Middle position Pulled out U TR30 2 100 for obser vation light...

Страница 33: ...ic observation tubes with normal field capable of the adjustments of the positioning angle 0 to 25 as well as front rear positioning by 45 mm of the eyepieces The U ETBI is the erect image model and t...

Страница 34: ...the center of the field of view 6 Gradually open the field iris diaphragm The condenser is properly centered if the iris image is centered and inscribed in the field of view 7 During actual use open...

Страница 35: ...n tube you will be able to see the objective pupil and aperture iris diaphragm image as shown in Fig 46 2 Compatibility of Objectives and Condensers Objective Magnification Condenser Achromat aplanat...

Страница 36: ...0 or higher the number applies only when oil is applied between the slide glass and the top surface of the condenser When oil is not present the NA is about 0 9 5 After use remove oil from the object...

Страница 37: ...he parfocality adjustment method refer to the instruction manual for the camera adapter in use 1 System Chart TV camera digital camera Camera mount adapter Camera adapter C mount camera adapter Trinoc...

Страница 38: ...not attached properly Re attach it 44 Revolving nosepiece is not attached properly Push the side dovetail all the way until it is stopped 44 An objective that falls outside of the condenser s illumina...

Страница 39: ...not properly centered Center the condenser 30 Condenser is lowered too far Adjust the condenser height position 30 Objective lowers too much and dest roys specimen Stage adapter is not installed prope...

Страница 40: ...U TBI 3 U TBI3 CLI U TR30 2 U TR30NIR U TTR 2 Widefield binocular Widefield tilting binocular Widefield trinocular Widefield tilting trinocular Field No 22 Tube inclination 30 5 35 continuously variab...

Страница 41: ...environment Indoor use Altitude Max 2000 meters Ambient temperature 5 to 40 C 41 to 104 F Maximum relative humidity 80 for temperatures up to 31 C 88 F decreasing linearly through 70 at 34 C 93 F 60...

Страница 42: ...nFLN Ph 4X Ph 10X2 Ph 20X Ph 40X Ph 40XO 60X 60XOI Ph 100XO2 Ph 100XOI2 0 13 0 3 0 5 0 75 1 3 0 9 0 65 1 25 1 30 0 6 1 30 17 0 10 0 2 1 0 51 0 2 0 2 0 12 0 2 0 2 0 17 0 17 0 17 0 17 0 17 0 17 0 17 2 5...

Страница 43: ...ic and the U MCZ controller requires the use of the provided Allen wrench for clamping the screws to ensure the performance we recommend that you have your Olympus representative install or uninstall...

Страница 44: ...llar using the provided clamping screws and Allen wrench Be sure not to mistake the left and right 2 After attaching the rear side of the stage on the microscope frame tighten the clamping screws on t...

Страница 45: ...mm attached with BX63 for tightening 3 Tighten the screw holes 3 and 4 with the clamping screw attached with BX63 from the lower side of the stage fixing pillar of BX63 Use Allen wrench 28x80mm attach...

Страница 46: ...tached with BX63 for tightening 3 Tighten the screw holes 2 and 3 with the clamping screw attached with BX63 from the lower side of the stage fixing pillar of BX63 Use Allen wrench 40x70mm attached wi...

Страница 47: ...h a soft cloth 5 Slide the lamp housing cover onto the housing base from the above Tighten the clamping screw while pressing downward on the cover Fig 49 Caution for Bulb Replacement During Use or Rig...

Страница 48: ...es and therefore cannot use eyepieces equipped with positioning pins When using an eyepiece with micrometer disk insert it in the right eyepiece sleeve When doing so make sure that the eye piece posit...

Страница 49: ...wer ON time has exceeded 20 000 hours 2 Lamp does not light sometimes even though the main switch is set to on 3 Illumination flickers when you move the lamp cable or lamp housing 4 Lamp cable is unus...

Страница 50: ...Setting 1 BF plan semi apo set Applicable observations BF PO FL BX3 UCD8A motorized universal condenser BX3 RFAA illuminator Top lens Turret No Turret No Nosepiece position Objective U TLD 1 2 3 4 5...

Страница 51: ...ator Top lens Turret No Turret No Nosepiece position Objective U TLD 1 2 3 4 5 6 7 8 Not moved PH DF BF DIC PO 1 PLN2X PLCN2X OUT BF PO Idle position PH DIC DF 7 Idle position 8 U FDICT 2 PLN4X PLCN4X...

Страница 52: ...lluminator Top lens Turret No Turret No Nosepiece position Objective U TLD 1 2 3 4 5 6 7 8 Not moved PH DF BF DIC PO 1 UPLFLN4X OUT BF PO Idle position DIC 7 Idle position 8 U FDICT 2 UPLFLN10X2PH IN...

Страница 53: ...ing Table 1 Certified Cord A power supply cord should be certified by one of the agencies listed in Table 1 or comprised of cordage marked with an agency marking per Table 1 or marked per Table 2 The...

Страница 54: ...Cables BASEC BASEC HAR 10 10 30 N V KEMA KEMA KEUR HAR 10 30 30 SEMKO AB Svenska Elektriska Materielkontrollanstalter SEMKO HAR 10 10 50 sterreichischer Verband f r Elektrotechnik VE VE HAR 30 10 50...

Страница 55: ......

Страница 56: ...many 3500 Corporate Parkway Center Valley Pennsylvania 18034 0610 U S A 491B River Valley Road 12 01 04 Valley Point Office Tower Singapore 248373 31 Gilby Road Mount Waverley VIC 3149 Melbourne Austr...