Graphical Digital Panel Meter WPMZ-3 INSTRUCTION MANUAL
94
IM-0883-01
Amplitude hold
Hold amplitude (difference between mzxmum and minimum values) of the section.
If plus/minus over occurs, it is not hold.
When recovering from plus over or minus over, the amplitude value is updated and held with the value immediately
after returning.
Deviation hold
Set reference value and hold display value that has the maximum absolute value of difference from reference value.
Caution, it holds not difference value but display value.
If plus/minus over occurs, it is not hold.
When recovering from plus over or minus over, the deviation value is updated and held with the value immediately
after returning.
You need to set reference value (DevBaseValue).
Sectikon state
Display value
On section
0
Amplitude hold
Section state
Display value
On section
0
Reference
value
Display value
Deviation hold