Graphical Digital Panel Meter WPMZ-3 INSTRUCTION MANUAL
32
IM-0883-01
DispUnit
Select a unit for display value.
Operation:[MENU]
⇒
[Input]
⇒
[ProcessInputA][ProcessInputB]
⇒
[DispUnit]
Setting value:μm,mm,cm,m,inch,g,kg,t,N,kN,MN,mN
・
m,N
・
m,kN
・
m,Pa,kPa,MPa,hPa,
kg/m2,N/m2,CustomUnit
Offset
Calibrate offset of input signal. Please calibrate both “Offset” and “Fullscale”.
●Operation for calibration (ex: InputRange 4-20mA, set display value to be 0-10000)
1. Select "Offset" and press "ENTER" key. (Go to offset setting screen)
2. Input 4mA to WPMZ and select "Read input val" and press "ENTER" key. Then, input current is
displayed in “Input”.
3. Set “Disp” to be “0”.
4. Back to forward screen by press “Left” key.
*You can also set “Input” manually.
Operation:[MENU]
⇒
[Input]
⇒
[ProcessInputA][ProcessInputB]
⇒
[Offset]
Setting value:-99999
~
99999 (DispValue)
*Setting values of “DecPoint” and “DispUnit” are reflected.
*If you select “CustomUnit”, you can create an original
unit within 6 letters. You can use following letters to
create a unit. ([a]
~
[z], [A]
~
[Z],
[,],(,),