Graphical Digital Panel Meter WPMZ-3 INSTRUCTION MANUAL
28
IM-0883-01
PatternSelect
You can store and set up 8 patterns (8 kinds) of various settings such as input and output with this unit, you can
select the pattern to save the setting in "PatternSelect".
The pattern number being set is displayed as "P5" on the upper right of the setting screen as shown below.
Operation:[MENU]
⇒
[Input]
⇒
[StrainInputA][StrainInputB]
⇒
[PatternSelect]
Setting value:Pattern1,Pattern 2,
・・・・・
Pattern 8
*At the time of opening the menu, the setting pattern is the value selected as operating on the measurement
screen.
* Since the setting pattern is common in each setting, "PatternSelect" such as output setting is also changed
when "PatternSelect" is changed by "StrainInput A" or "StrainInput B".
BridgeExcitation
Select bridge power supply voltage for sensor.
Please close the menu once for reflecting the setting before connect a sensor.
Operation:[MENU]
⇒
[Input]
⇒
[StrainInputA][StrainInputB]
⇒
[BridgeExcitation]
Setting value:2.5V,5V(Default),10V
*The maximum supply current is 30mA
@
2.5V, 60mA
@
5V, 30mA
@
10V
DecPoint
Select decimal point position of display value.
Operation:[MENU]
⇒
[Input]
⇒
[StrainInputA][StrainInputB]
⇒
[DecPoint]
Setting value:##### (Default),####.#,###.##,##.###,#.####
DispUnit
Select a unit for display value.
Operation:[MENU]
⇒
[Input]
⇒
[StrainInputA][StrainInputB]
⇒
[DispUnit]
Setting value:μm,mm,cm,m,inch,g,kg,t,N,kN,MN,mN
・
m,N
・
m,kN
・
m,Pa,kPa,MPa,hPa,
kg/m2,N/m2,CustomUnit
*If you select “CustomUnit”, you can create an original
unit within 6 letters. You can use following letters to
create a unit. ([a]
~
[z], [A]
~
[Z],
[,],(,),
1
,
2
,
3
,
1
,
2
,
3
,
-
,μ,Ω,g,
・
,/,ℓ,%,%
0
,°,’,” )