Graphical Digital Panel Meter WPMZ-3 INSTRUCTION MANUAL
105
IM-0883-01
DifMag
Set difference magnification of maximal and minimal.
*This setting is enable if “HoldType” is “DifferenceHold”.
Operation:[MENU]
⇒
[Input]
⇒
[MultiA/B S1/2/S3/S4]
⇒
[DifMag]
Setting Value: 0.01
~
99.99[Times] (Initial state is 1.00)
InfTimeA
Set time of “InfTimeA” for inflection hold.
*This setting is enable if “HoldType” is “InflectionHold”.
Operation:[MENU]
⇒
[Input]
⇒
[MultiA/B S1/2/S3/S4]
⇒
[InfTimeA]
Setting value: 1
~
499[points] (Initial state is 100)
*Please be aware that the actual time varies depending on sampling rate.
Example) Sampling rate:2000
回
/sec, InfTimeA:200
Actual inflection = (1/2000) × 200 = 0.1sec
InfTimeB
Set time of “InfTimeB” for inflection hold.
*This setting is enable if “HoldType” is “InflectionHold”.
Operation:[MENU]
⇒
[Input]
⇒
[MultiA/B S1/2/S3/S4]
⇒
[InfTimeB]
Setting value: 1
~
499[points] (Initial state is 1)
*Please be aware that the actual time varies depending on sampling rate.
Example) Sampling rate:2000
回
/sec, InfTimeA:200
Actual inflection = (1/2000) × 200 = 0.1sec
InfValueZ
Set value of “InfValueZ” for inflection hold.
*This setting is enable if “HoldType” is “InflectionHold”.
Operation:[MENU]
⇒
[Input]
⇒
[MultiA/B S1/2/S3/S4]
⇒
[InfValueZ]
Setting: ±99999 (Initial state is 1000)
CompOutput
Set output that will output if alarm of this section turns on.
Operation:[MENU]
⇒
[Input]
⇒
[MultiA/B S1/2/S3/S4]
⇒
[CompOutput]
Setting: None (Initial state), AL1, AL2, AL3, AL4
CompAlarmCond.
Select a condition of comparison alarm.
Outside:
Alarm turns on if hold value is outside of the zone.
Inside:
Alarm turns on if hold value is inside of the zone.
Operation:[MENU]
⇒
[Input]
⇒
[MultiA/B S1/2/S3/S4]
⇒
[CompAlarmCond.]
Setting: OutsideTheZone (Initial state), InsideTheZone