Graphical Digital Panel Meter WPMZ-3 INSTRUCTION MANUAL
29
IM-0883-01
AutoAdjust
Do actual load calibration.
As shown in the figure below, it has 2 screen configurations, and "Zero point calibration" "Actual load
calibration" is continued and executed.
●Operation for calibration
1. Connect the sensor and set it to no load condition.
2. Select "Execute" and press "ENTER" key.
3. Select "Next" and press "ENTER" key.
4. Apply actual load to the sensor.
5. Set the input load value to "SetValue".
6. Select "Execute" and press "ENTER" key.
Operation:[MENU]
⇒
[Input]
⇒
[StrainInputA][StrainInputB]
⇒
[AutoAdjust]
Setting value:-99999
~
99999 (SetValue)
*Setting values of “DecPoint” and “DispUnit” are reflected.
ManuAdjust
Do equivalent load calibration.
As shown in the figure below, it has 2 screen configurations, and "Zero point calibration" "Equivalent load
calibration" is continued and executed.
●Operation for calibration
1. Connect the sensor and set it to no load condition.
2. Select "Execute" and press "ENTER" key.
3. Select "Next" and press "ENTER" key.
4. Set rated output value of sensor to “Rating”.
5. Set rated capacity value of sensor to “SetValue”.
6. Select "Execute" and press "ENTER" key.
Operation:[MENU]
⇒
[Input]
⇒
[StrainInputA][StrainInputB]
⇒
[ManuAdjust]
Setting value:-99999
~
99999 (Rating), -99999
~
99999 (SetValue)
*Setting values of “DecPoint” and “DispUnit” are reflected.