Graphical Digital Panel Meter WPMZ-3 INSTRUCTION MANUAL
100
IM-0883-01
How to measure of multi hold function
How to measure of multi hold function is shown below.
Measurement start is executed by MultiA function if you measure Ach, and MultiB function if you measure Bch that
are operated by external control, shortcut key and communication function.
①
[MultiHoldA]
Hold start condition
②
Waiting for hold start condition of Section 1
③
Section 1
④
Interval between section 1 and 2
※Skip to ⑤ if the SectionSwitch is
“LevelMethod” or “EdgeMethod”
※Display immediately before hold value.
If no hold detected, it free-runs.
⑤
※“NG” and latched output continues.
⑥
⑦
Result
Clear all warning outputs.
Go to ②
※No interval exist after
section 4, and display result.
※It free-runs until next measurement start.
Abbreviation
Display ”READY”
Display “WAIT”
Display “RUN”
Display “END”
Display “FIN.”
Display “RUN”
※“NoDetect” will be displayed if there is
no hold in the section.
You can choose “NoDetect” to alarm or
not alarm with setting.(It is alarm by default)
※
will light up if there are no alarms in all sections.
Waiting for
External control
[MultiHoldA]
※Skip to ③ if the HoldStartCond. is “Normal”
Waiting for hold start condition of Section 2
※Skip if the HoldStartCond. is “Normal”
Section 4
[MultiHoldA]
※“NG” lights up and warning output
turns ON and latched if there is an alarm.
“NG” and latched output continue until the end of ⑦.