Graphical Digital Panel Meter WPMZ-3 INSTRUCTION MANUAL
54
IM-0883-01
NormalHold: As long as the external control terminal is set to the L level, the display continues to be updated
with the most distant value. By setting the pin to H level, holding is canceled and the latest display
value is displayed.
AreaHold: As long as the external control terminal is set to the L level, the display continues to be
updated with the most distant value. When the terminal is set to H level, it keeps holding
at the the most distant value at that point.
AveHold
This function performs additional moving average while running.
The average number of times is set by the “AveHoldCount” of the setting item.
It is effective in temporarily canceling vibration influences etc. by the external environment.
There are two modes of “NormalHold” and “AreaHold”, and the mode can be changed by the “HoldMode”.
NormalHold: As long as the external control terminal is set to the L level, the display continues to be
updated with the moving average value. By setting the pin to H level, holding is canceled
and the latest display value is displayed.
AreaHold: As long as the external control terminal is set to the L level, the display continues to be
updated with the moving average value. When the terminal is set to H level, it keeps holding
at the moving average value at that point.
HoldReset
This is a function to cancel all holding state and display the latest display value during function ON.
In addition, when the measurement mode is “Multi hold”, if holding reset is executed during operation of
sections 1 to 4,
ends all sections immediately and moves to the result sequence.
While the function is active, the result sequence will be maintained and the next section will not be started.
For detailed operation of multi hold, please refer to the section on multi hold.
The function becomes valid while the terminal is "0" level or shorted with the COM terminal.
DigitalZero
The digital zero function (DZ) is a function to set the display value to zero when the DZ instruction is made. The
display is offset during the DZ instruction, and the fluctuation width from the start of instruction is treated as
the display value.
"DZ" icon is displayed on the measurement screen during operation.
Display value
0
DZ: OFF
DZ: ON
External control
0
V
Short with COM
or
0 level
Time
Display value
Sensor value
On the section