Graphical Digital Panel Meter WPMZ-3 INSTRUCTION MANUAL
59
IM-0883-01
There are three operation methods of external control function, terminal control, shortcut, and communication
function, but there are the relationships shown in the table below as the operation at the time of simultaneous
execution.
Control function list
Shows the list of functions that can be registered as a shortcut function.
Please check each function description for each operation details.
"ManuAdjust" "AutoAdjust"
⇒
Details are in “8-1. CALIBRATION OF STRAINGAUGE INPUT”
Other control functions
⇒
describe details in “11. EXTERNAL CONTROL FUNCTION”.
機能
動作
None
None
CompareReset
Function to turn off all comparison result and its output.
DispHold A/B/A&B
Function to hold display value of current value.
MaxHold A/B/A&B
Function to hold the maximum value of display.
MinHold A/B/A&B
Function to hold the minimum value of display.
AmpHold A/B/A&B
Function to hold difference between maximum and minimum.
DevHold A/B/A&B
Function to hold a display value most distant from
an arbitrary reference value.
AveHold A/B/A&B
Function to stabilize the display by performing additional moving
average for the specified number of times.
HoldResetA/B/A&B
Function to reset holding state.
DigitalZero A/B/A&B
Function to zero display value.
TrendLog
Function to start trend logging, invalid if logging is already in
progress.
PatternChange 1/2/3
Function to change active pattern (up to 3 terminals used).
WaveCompare A/B/A&B Instruction to start / stop waveform comparison operation.
MultiHold A/B
Perform multi-hold section control.
ManuAdjust A/B
Open calibration screen for actual load calibration.
AutoAdjust A/B
Open calibration screen for equivalent load calibration.
*A/B/A&B in the function column indicate Ach operation, Bch operation, Ach / Bch simultaneous
operation, respectively.
* Only pattern change function, pattern is switched in order shown below. Pattern changing is
carried out every seconds when you press and hold the button.
Pattern 1
→
2
→
3
→
4
→
5
→
6
→
7
→
8
→
1
→
(continue)
* Each holding function, waveform comparison, multi hold, and trend log are displayed only
in the corresponding measurement mode.
* Actual load calibration and equivalent load calibration are displayed only when the target input channel
is straingauge input.
ON operation If one of the control methods is ON, the function will be ON.
OFF
operation
The function is canceled when both control states (terminal control and communication)
are OFF.
The function ON state by shortcut is automatically canceled by OFF operation either of
terminal control or communication.
Trigger
operation
It is possible to operate simultaneously by any control method.
However, if each function is not ready for operation again, it can not operate.