DelayTimer
Hold starts after certain time after section starts.
You need to set more settings below if [HoldStartCond.] is “DelayTimer”.
・
DelayTimer:
Set delay time.
Comparison judgement
About comparison judgement is mentioned below.
Comparison judgement is activated with hold value and comparison judgment value.
You can select always compare in section or compare once at end of section.
The comparison judgment values are set in the upper and lower limit zones, and it can be selected whether it is
alarm when the hold value is outside the zone or alarm when inside the zone.
These settings are set per section.
CompTiming
Select always compare in section or compare once at end of section.
Default setting is always comparing in section.
CompJudgementValue
Set upper and lower limits of comparison judgement zone.
*If you want to use one of both, set the other value to +99999 or -99999.
CompAlarmCond.
Select whether it is alarm when the hold value is outside the zone or alarm when inside the zone.
By default, it is alarm when the hold value is outside the zone.
CompOutput
Select alarms which alarm. By default, no alarm is selected.
You can select same alarm which is selected at other section setting. Then, it alarms either sections.