Graphical Digital Panel Meter WPMZ-3 INSTRUCTION MANUAL
33
IM-0883-01
Fullscale
Calibrate fullscale of input signal. Please calibrate both “Offset” and “Fullscale”.
●Operation for calibration (ex: InputRange 4-20mA, set display value to be 0-10000)
1. Select "Fullscale" and press "ENTER" key. (Go to fullscale setting screen)
2. Input 20mA to WPMZ and select "Read inpu val" and press "ENTER" key. Then, input current is
displayed in “Input”.
3. Set “Disp” to “10000”.
4. Back to forward screen by press “Left” key.
*You can also set “Input” manually.
Operation:[MENU]
⇒
[Input]
⇒
[ProcessInputA][ProcessInputB]
⇒
[FullScale]
Setting value:-99999
~
99999 (DispValue)
*Setting values of “DecPoint” and “DispUnit” are reflected.
2 INPUT CALCULATION
*In 1ch products, this setting item is invisible.
Setting of 2 input calculation is proceed as follow.
Each setting items are set up at [Setting]
⇒
[Input]
⇒
[2InputCalc]
Descriptions of each setting item are described on the following pages after this section.
Moreover, please check "input detailed setting" for the setting not described in the above flow.
*1 Each setting and calibration value are saved and reflected at the timing when the menu is closed and
returned to
the measurement screen.
Note that the contents will not be saved if the power is turned off before closing the menu.
Pattern select
Select arithmeric
expresion
Select arithmeric
coefficient
Decimal point
position select
Unit select