Graphical Digital Panel Meter WPMZ-3 INSTRUCTION MANUAL
85
IM-0883-01
18.
MULTI HOLD MODE
Overview
In the multi hold mode, measurements are divided into four sections of section 1 to section 4, arbitrary hold settings
are set to each section, and comparison output is performed by hold value in each section and comparison judgment
value set for each section, respectively.
To set this mode, operate[MENU]
⇒
[System]
⇒
[General]
⇒
[MeasureMode]
⇒
[MultiHold].
How to use of multi hold mode is shown below.
Each settings are set into [MENU]
⇒
[Inputm]
⇒
[MultiA/B Base] , [MultiA/B S1
~
S4].
SectionSwitch
LevelMethod / EdgeMethod /
EdgeTimer / AutoTimer
Set parameters for
each hold type.
CompOutput
HoldType
*You can set a hold type to each sections. Hold types are below.
[None], [PeakHold], [BottomHold], [AmplitudeHold],
[DevHold], [MaximalHold], [MinimalHold], [DifferenceHold],
[InflectionHold]
Measurement
HoldStartCond.
Normal / Threshold /
StartDelay
*You need to set settings below for each sections.
CompOutput: Select one of None / AL1~AL4
CompAlarmCond.: Select Outside / Inside
CompJudgeValue: Set upper limit/lower limit
CompTiming: Compare judgement is executed
when the section finished or
compare judgement is always
execute.
* In this mode, instruct measurement start and section change
by executing "MultiHoldA" and "MultiHoldB" function by
"external control", "shortcut key" or "communication".
The measurement is interrupted by “HoldReset” function.
*You need to set each settings below if you select “Threshold”.
・
Threshold
・
ThresholdDir
・
ThresholdTimeout
・
ThresholdTimeoutOutput
*You need to set setting below if you select “StartDelay”.
・
DelayTimer
CompAlarmCond.
CompJudgeValue
CompTiming
*You need to set each settings below if you select
“EdgeTimer” or “AutoTimer”.
・
SectionTimerS1 (Set time of section 1.)
・
SectionTimerS2
・
SectionTimerS3
・
SectionTimerS4
*If hold type is [None], [PeakHold], [BottomHold] or
[AmplitudeHold], you need to set no parameters.
Setting for comparison judgement