Graphical Digital Panel Meter WPMZ-3 INSTRUCTION MANUAL
104
IM-0883-01
ThresholdTimeout
When the “HoldStartCond.” is “Threshold”, if the hold start condition is not satisfied for a certain time, this is a
function to end the section as "No Detect".
Set the threshold timeout time.
*This setting is enable if “HoldStartCond.” is “Threshold”.
*Set “TresholdTimeoutOutput” together.
Operation:[MENU]
⇒
[Input]
⇒
[MultiA/B S1/2/S3/S4]
⇒
[ThresholdTimeout]
Setting value: 0.00
~
99.99[sec] (Initial state is 0.00)
*This function is disable when the setting value is 0.00[sec].
ThresholdTimeoutOutput
Set alarm output that is output when threshold timeout occurred.
*This setting is enable if “HoldStartCond.” is “Threshold”.
*This output is reset when the next multi hold started.
Operation:[MENU]
⇒
[Input]
⇒
[MultiA/B S1/2/S3/S4]
⇒
[ThresholdTimeoutOutput]
Setting: None (Initial state), AL1, AL2, AL3, AL4
DelayTimer
Set the delay time.
*This setting is enable if “HoldStartCond.” is “DelayTimer”.
Operation:[MENU]
⇒
[Input]
⇒
[MultiA/B S1/2/S3/S4]
⇒
[DelayTimer]
Setting: 0.01
~
99.99[sec] (Initial state is 0.01)
HoldType
Select hold type of section.
Operation:[MENU]
⇒
[Input]
⇒
[MultiA/B S1/2/S3/S4]
⇒
[HoldType]
Setting: None (Initial state), PeakHold, BottomHold, AmplitudeHold, DevHold,
MaximaalHold, MinimalHold, DifferenceHold, InflectionHold
DevBaseValue
Set a reference value of deviation hold.
*This setting is enable if “HoldType” is “DevHold”.
Operation:[MENU]
⇒
[Input]
⇒
[MultiA/B S1/2/S3/S4]
⇒
[DevHold]
Setting value: ±99999 (Initial state is 10000)
DifValue
Set reference value of difference hold.
*This setting is enable if “HoldType” is “DifferenceHold”.
Operation:[MENU]
⇒
[Input]
⇒
[MultiA/B S1/2/S3/S4]
⇒
[DifValue]
Setting value: 99999 (Initial state is 1000)