Graphical Digital Panel Meter WPMZ-3 INSTRUCTION MANUAL
44
IM-0883-01
Operation:[MENU]
⇒
[Input]
⇒
[StrainInputA][StrainInputB][ProcessInputA][ProcessInputB]
⇒
[TrackingZero]
Setting value:
Interval 0 to 9999[msec](Default is 0)
ActiveArea
±99999(Default is 0)
* If either of “Interval” or “ActiveArea” is 0, the tracking zero function is invalid.
* As the power is turned off, the offset value corrected by tracking zero becomes 0.
DispLimit
Sets the upper and lower limit of the display value. It is effective when you want to suppress output to 10 V even
if input is over (ex: 22mA) when input is 4 to 20 mA and output is 0 to 10 V.
ex) When the display value 0 to 10000 is set for the input 4-20 mA and the analog output is 0 to 10 V, the analog
output becomes maximum 10 V by setting the “UpperLimit” 10000. (The output is limited to 10 V even when an
excessive input of 22 mA is entered)
Operation:[MENU]
⇒
[Input]
⇒
[StrainInputA][StrainInputB][ProcessInputA][ProcessInputB]
⇒
[DispLimit]
Setting value:
LowerLimit ±99999 (Default is -99999)
UpperLimit
±99999(Default is +99999)
DispLoCut
It is a function to cut floating input near zero. If the absolute value of the display value is less than the set value,
the display value is set to zero.
Operation:[MENU]
⇒
[Input]
⇒
[StrainInputA][StrainInputB][ProcessInputA][ProcessInputB]
⇒
[DispLoCut]
Setting value: 0
~
99999 (Default is 0)
* Calculation processing of “DispLoCut” is performed after calculation processing of digital zero function.
InsDispStep
When the minimum digit of display value fluctuates, truncate the minimum digit and stabilize the display.
Operation:[MENU]
⇒
[Input]
⇒
[StrainInputA][StrainInputB][ProcessInputA][ProcessInputB]
⇒
[InsDispStep]
Setting value: None (Default): 0 to 9
5steps
:
0 to 4 → 0,
5 to 9 → 5
10steps :
0 to 9 → 0
ZeroArea
Zero detection is a function to judge whether the display value is around zero and disable comparison output
function when it is near zero.
It is effective when measuring the weight of an object, such as suppressing comparison ountput function when
not holding anything on the tare.
In this item, set the range of the display value considered to be near zero.
* To enable this function, it is necessary to set the “ActCondition” of the comparison output function.
(The setting method of the “ActCondition” is [MENU]
⇒
[Output]
⇒
[CompareAL1
~
4]
⇒
[ActCondition]
Operation:[MENU]
⇒
[Input]
⇒
[StrainInputA][StrainInputB][ProcessInputA][ProcessInputB]
⇒
[ZeroArea]
Setting value: 0
~
99999 (Default is 0)