Graphical Digital Panel Meter WPMZ-3 INSTRUCTION MANUAL
56
IM-0883-01
WaveCompare
This is a function to start or stop measurement when “MeasureMode” is “WaveCompare”.
This function is invisible when “MeasureMode” is “Default” or “MultiHold”.
This function is executed by shorting the terminal for 20 ms or more or setting it to "0" level (falling edge
detection).
Please check [17. WAVE COMPARE MODE].
*Basically, it functions as a capture start indication, but when detecting a falling edge again during the detection
wait state and capture operation, capture stops.
MultiHold
This is a function to switch sections in “Multihold” mode.
This function is invisible when “MeasureMode” is “Default” or “WaveComp”.
This function is executed by shorting the terminal for 20 ms or more or setting it to "0" level (falling edge
detection).
Please check [18. MULTI HOLD MODE].
Setting Items
DispMode
You can switch the operating mode of DispHold either “FreeRun” or “OneShot”.
FreeRun: The display value is held while the external control terminal is set to low level.
OneShot: The display value is kept constant all the time, and it is updated every time the external control
terminal falls.
For details, please refer the content about “DispHold”.
Operation: [MENU]
⇒
[Input]
⇒
[HoldA] [HoldB]
⇒
[DispMode]
Setting value: FreeRun (Default), OneShot
DispDelay
You can set the delay about when the DispHold function is canceled.
There are differences in the delay operation for each “FreeRun” and “OneShot” shown below.
“FreeRun”: You can set the time from release of device to release of hold.
“OneShot”: You can set the time from release of device to update of display value.
This function is disable when the setting value is 0.
Operation: [MENU]
⇒
[Input]
⇒
[HoldA] [HoldB]
⇒
[DispDelay]
Setting value:0
~
99.99[sec]
(Default is 0)
CAUTION
Setting of retention cancellation delay is effective only for "DispHold".
No delay occurs when releasing other holding functions.
HoldMode
You can switch the operation mode of each holding function except "DispHold" function to "NormalHold" or
"AreaHold".
The basic operation of each mode is as follows.
"NormalHold": The hold function is executed while the terminal is short-circuited and released when the
terminal is released
"AreaHold": The hold function is executed while the terminal is short-circuited, and the display value is fixed
with the final
holding value when the terminal is released
For details of operation of each holding function, please confirm the holding function explanation.