Graphical Digital Panel Meter WPMZ-3 INSTRUCTION MANUAL
38
IM-0883-01
PatternSelect
You can store and set up 8 patterns (8 kinds) of various settings such as input and output with this unit, you can
select the pattern to save the setting in "PatternSelect".
The pattern number being set is displayed as "P5" on the upper right of the setting screen as shown below.
Operation:[MENU]
⇒
[Output]
⇒
[CompareAL1], [CompareAL2], [CompareAL3], [CompareAL4]
⇒
[PatternSelect]
Setting value:Pattern1, Pattern 2,
・・・・・
Pattern 8
* At the time of opening the menu, the setting pattern is the value selected as operating on the measurement
screen.
* Since the setting pattern is common in each setting, "PatternSelect" such as input setting is also changed when
"PatternSelect" is changed by "CompareAL1
~
4”.
OutputDispValue
Select reference source of comparison output. You can set AL1~AL4 arbitrarily to Ach, Bch or Calc.
* Even if the comparison output is allocated to a display element that is not displayed on the measurement
screen, it will monitor and output when the comparison result turns ON.
*In 1ch products, [Bch] and [Calc] are invisible.
Operation:[MENU]
⇒
[Output]
⇒
[CompareAL1], [CompareAL2], [CompareAL3], [CompareAL4]
⇒
[OutputDispValue]
Setting value: None(Default),Ach,Bch,Calc
CompareMode
Select compare mode: Leveljudge or Zonejudge or Diffjudge.
Compare mode
Function
Leveljudge
[Threshold] is set, and it is judged whether the displayed value is less than (or
exceeded) the judgment value.
Zonejudge
[ZoneLowerLimit] and [ZoneUpperLimit] are set, and it is judged whether or not
the displayed value is within (or out of) the zone area.
Diffjudge
The maximum value and the minimum value are monitored in
[DetectionInterval], and when the change amount (maximum value - minimum
value) becomes excess [ChangeAmount], the comparison output is output ON.
Operation: [MENU]
⇒
[Output]
⇒
[CompareAL1], [CompareAL2], [CompareAL3], [CompareAL4]
⇒
[CompareMode]
Setting value: Leveljudge(Default), Zonejudge, Diffjudge