Graphical Digital Panel Meter WPMZ-3 INSTRUCTION MANUAL
119
IM-0883-01
SETTING VARIABLES
1st Layer
2nd Layer
3rd Layer
(setting items)
4th Layer (setting values)
Remarks
Initial values
Settable values
1.Input
ProcessInput
A
ProcessInput
B
StrainInputA
StrainInputB
PatternSelect
Pattern in use
Pattern 1/2/3/4/5/6/7/8
Select pattern No. to set.
InputRange
±3.5mV/V
(0
~
10V)
±3.5mV (0
~
5V/1
~
5V/±5V/0
~
10V
/±10V/0
~
20mA /4
~
20mA/±20mA)
Select an input range. Settings of process
input are in ().
*In process input, offset and fullscale
setting are initialized by changing this
item.
SensorPower
12V
12V/24V
Only in process input
BridgeExcitation
5V
2.5V/5V/10V
Only in straingauge input
AnalogFilter
300Hz
OFF/30Hz/300Hz/600Hz
Only in straingauge input
Sampling rate
4000times/sec
(1 input product)
2000times/sec
(2 input product)
4000 times/sec, 2000, 1000, 500, 200, 100,
20,
10, 5, 2, 1,
Once per 2sec, Once per 5sec, Once per
10sec
Select sampling rate
(4000times/sec is enable only 1ch
product)
MoveAve
32
None/2/4/8/16/32/64times
Set moneing average of input.
Offset
0
Go to screen of setting offset.
Only process input.
Fullscale
10000
Go to screen of setting fullscale.
Only process input.
ManuAdjust
0
~
35000
Go to screen of manual adjust.
Only straingauge input.
AutoAdjust
0
~
35000
Go to screen of auto adjust.
Only straingauge input.
DecPoint
#####
#####/ #.####/ ##.###/ ###.##/ ####.#
Initial state depends on measurement
range.
DispUnit
None
None/Pa, kg, N, etc./CustomUnit
InputCorrect
None
None/Linearize
LinearizePoint
1stInput:
0
1stOutput:
0
2ndInput:1000
2ndOutput:1000
……
16thInput:15000
16thOutput:15000
Each: ±99999
Enable only if [InputCorrect] is
[Linearize].
DispShift
0
±99999
Take over setting of [DecPoint].
TrackingZero
Interval:
0
ActiveArea:
0
Interval:
0
~
99.99[sec]
ActiveArea:
0
~
99999
Disable if setting value is 0[sec].
Take over setting of [DecPoint].
DispLimit
LowerLimit:
-99999
UpperLimit:
+99999
±99999
±99999
Take over setting of [DecPoint].
DispLoCut
0
0
~
99999
Cut both positive and negative.
InsDispStep
None
None/5steps/10steps
ZeroArea
0
0
~
99999[digit]
Disable if the value is 0[digit].
StableArea
0
0
~
99999[digit]
Disable if the value is 0[digit].
StableTime
0
0
~
99.99[sec]
Disable if the value is 0[sec].
2InputCalc
PatternSelect
Pattern in use
Pattern 1/2/3/4/5/6/7/8
Select pattern No. to set.
Expression
None
None/Add((A+B)+C)*K /Sub((B-A)+C)*K
/Mul((A*B)+C)*K /Div((B/A)+C)*K
/Ave (((A+B)/2)+C)*K
/HiSelect((Larger of A and B)+C)*K
/LoSelect((Smaller of A and B)+C)*K
/DifAbs(Abs of (B-A))+C)*K
/ErrRaito((A/B)-1)*K /Dens(B/(A+B))*K
/Add(A+B)*K+C /Sub(B-A)*K+C
/Mul(A*B)*K+C /Div(B/A)*K*C
Select a expression.
Const-C
C: 0
C: ±9.9999 x 10^(±5)
Coef-K
K: 1
K: ±9.9999
DecPoint
#####
#####/ #.####/ ##.###/ ###.##/ ####.#
Initial state depends on measurement
range.
DispUnit
None
None/Pa, kg, N, etc./CustomUnit
InsDispStep
None
None/5steps/10steps
DispLimit
LowerLimit:
-99999
UpperLimit:
+99999
±99999
±99999
Take over setting of [DecPoint].
ExternalCtrl
ExtCtrl 1
~
5 Func
None
None/CompareReset/HoldResetA,B,A&B
/DispHoldA,B,A&B/Peak_HoldA,B,A&B
/BottomHoldA,B,A&B/AmpHoldA,B,A&B
/DevHoldA,B,A&B/AveHoldA,B,A&B
/DigitalZeroA,B,A&B/PatternChange1,2,3
/DispChange/TrendLog/WaveCompA,B
/MultiHoldA,B
Select a function which is attached to
each external control terminals.
Hold A/B
PatternSelect
Pattern in use
Pattern 1/2/3/4/5/6/7/8
Select pattern No. to set.
DispHoldMode
Normal
Normal/OneShot
HoldOffDelay
0.00
0.00
~
99.99[sec]
HoldMode
NormalHold
NormalHold/AreaHold
Select a mode of DispHold.
DevBaseValue
0
±99999
Take over setting of [DecPoint].
AveHoldCount
None
None/2/4/8/16/32/64times