Graphical Digital Panel Meter WPMZ-3 INSTRUCTION MANUAL
45
IM-0883-01
StableArea
It judges whether the display value is stable or not and invalidates the comparison output function when it is not
stable. If the variation of the display value at each specified time is within the set value, it is judged stable and
the “ST” mark will light up at the upper right of the display value on the measurement screen.
It is effective when measuring the weight of the object caught with a crane when you want to make a comparison
output function after the shake falls.
In this item, set the fluctuation range of the display value considered stable.
* Please set it together with “StableTime”. This function is invalid if either set value is 0.
* To enable this function, it is necessary to set the “ActCondition” of the comparison output function.
(The setting method of the “ActCondition” is [MENU]
⇒
[Output]
⇒
[CompareAL1
~
4]
⇒
[ActCondition]
Operation:[MENU]
⇒
[Input]
⇒
[StrainInputA][StrainInputB][ProcessInputA][ProcessInputB]
⇒
[StableArea]
Setting value: 0
~
99999 (Default is 0)
StableTime
Same function as “StableArea” and this item is a time parameter for detecting stability of display value.
* Please set it together with “StableArea”. This function is invalid if either set value is 0.
Operation:[MENU]
⇒
[Input]
⇒
[StrainInputA][StrainInputB][ProcessInputA][ProcessInputB]
⇒
[StableTime]
Setting value: 0
~
9999 * 10msec (Default is 0)
*You can set by 10ms steps.
2InputCalc
InsDispStep
When the minimum digit of display value fluctuates, truncate the minimum digit and stabilize the display.
Operation:[MENU]
⇒
[Input]
⇒
[2InputCalc]
⇒
[InsDispStep]
Setting value: None (Default): 0 to 9
5steps
:
0 to 4 → 0,
5 to 9 → 5
10steps :
0 to 9 → 0
DispLimit
Sets the upper and lower limit of the display value. It is effective when you want to suppress output to 10 V even
if input is over (ex: 22mA) when input is 4 to 20 mA and output is 0 to 10 V.
ex) When the display value 0 to 10000 is set for the input 4-20 mA and the analog output is 0 to 10 V, the analog
output becomes maximum 10 V by setting the “UpperLimit” 10000. (The output is limited to 10 V even when an
excessive input of 22 mA is entered)
Operation:[MENU]
⇒
[Input]
⇒
[2InputCalc]
⇒
[DispLimit]
Setting value:
LowerLimit ±99999 (Default is -99999)
UpperLimit
±99999(Default is +99999)