Graphical Digital Panel Meter WPMZ-3 INSTRUCTION MANUAL
27
IM-0883-01
8.
BESIC SETTINGS
CALIBRATION OF STRAINGAUGE INPUT
Calibration of straingeage input is proceed as follow.
Each setting items are set up at [Setting]
⇒
[Input]
⇒
[StrainInputA] or [StrainInputB].
When registering calibration values of multiple sensors, change the pattern and continue setting and
calibration.
Descriptions of each setting item are described on the following pages after this section.
Moreover, please check "input detailed setting" for the setting not described in the above flow.
*1 Each setting and calibration value are saved and reflected at the timing when the menu is closed and
returned to the measurement screen.
Note that the contents will not be saved if the power is turned off before closing the menu.
Pattern select
Bridge power supply
voltage select
Zero point
adjustment
Actual load
calibration
Equivalent
load calibration
Zero point
adjustment
Decimal point
position select
Unit select
Measurement start
*If you changed the bridge power supply, please
close the menu once for reflecting the setting.
Connect a sensor
*When actual loading to the sensor is possible,
please perform "Actual load calibration", and is
inpossible, please perform "Equivalent load
calibration".
"Zero point adjustment" is performed within
each calibration setting item.
*For wiring, please confirm
"5. CONNECTING TERMINALS”.