Graphical Digital Panel Meter WPMZ-3 INSTRUCTION MANUAL
83
IM-0883-01
RefWaveCapture
Create reference waveform for making comparison waveform by measuring waveform.
Measure waveforms as many times as specified, and create reference waveform by average the measured data.
Operation:[MENU]
⇒
[Input]
⇒
[WaveCompareA/B]
⇒
[RefWaveCapture]
Setting value:
Average 1
~
99
CompWavePos
Select which comparison waveform is used for judgement.
UpperAndLower: Set both upper and lower comparison waveforms for zone judgement.
UpperOnly: In order to judge whether or not it has exceeded, only the comparison waveform on the upper side is
set.
LowerOnly: In order to judge whether or not it has less than, only the comparison waveform on the lower side is
set.
Operation:[MENU]
⇒
[Input]
⇒
[WaveCompareA/B]
⇒
[CompWavePos]
Setting:
UpperAndLower(Initial state), UpperOnly, LowerOnly
CreateCompWave
Create comparison waveform from reference waveform.
Specify the vertical shift amount and the left / right shift amount, execute confirmation each time and create a
judgment waveform.
Operation:[MENU]
⇒
[Input]
⇒
[WaveCompareA/B]
⇒
[CreateCompWave]
Setting value: UD Shift 0
~
99999[digit]
LR Shift
0
~
99[
×
10points]
Check
⇒
You can check the comparison waveform.
AutoScale
Select autoscale enable/disable at [ChechCompWave] screen.
Autoscale function adjust screen scale to have 10% upper and lower margin.
Operation:[MENU]
⇒
[Input]
⇒
[WaveCompareA/B]
⇒
[AutoScale]
Setting
Enable (Initial state), Disable
Scale
Set scale of vertical for [ChechCompWave] screen.
*This function is enable only when [AutoScale] is “Disable”.
Operation:[MENU]
⇒
[Input]
⇒
[WaveCompareA/B]
⇒
[Scale]
Setting value
LowerLimit
±99999 (Initial state is 0)
UpperLimit
±99999 (Initial state is 10000)