Operation,
Screen Layout and Navigation
1-2-3
RGS-2000NG TCAS Test Set Operation Manual
139187 / Rev. 500
Page 182
Subject to Export Control. See Cover Page for details.
September 2021
F i g u r e 1 .2 . 3 - 1 6 2 U A T Sc e n a r i o S c re e n a n d S o f t k e y M e n u S e q u en c e
Screen Components
Description
Scenario Time Field
Defines the total Scenario Run Time.
Run Time
Displays the length of time a Scenario has been running.
UAT#1 / UAT #2
Dynamic Targets
Quantity: Defines the number of dynamic targets.
Enables: Displays the quantity of dynamic targets that are enabled.
Static Targets
Quantity: Defines the number of static targets.
Enables:Displays the quantity of static targets that are enabled.
Total Messages
Displays the total number of messages.
Enable
Enables/Disables sweep mode.
Sweep
Step: Defines the milliseconds time between swept transmissions.
Range 50 to 200 ms. Used only when sweep is enabled.
Interval: Defines the length or end point of the sweep in milli seconds.
Used only when sweep is enabled.
Capture Squitters and Data
Logging
Enables/disables the Capture Squitters and Data Logging. If enabled,
the log file is cleared when the scenario starts and the log file starts
capturing new messages.
Static Test Mode
Enables/disables the Static Test Mode. Targets are active at the end of
the scenario time at the target’s last position.
Re-Compile After Load
Enables/disables the Re-Compile After Load. If enabled, compiles all
messages for the different targets after loading a saved file.
UAT Test Mode
Selects the UAT Test Mode.
(Refer to the Definition of UAT Test Mode Table.)
UAT MSO Step
Defines the UAT MSO Step, the space between messages in MSO steps
(MSO Step = 250us). Only used for multiple targets.