Operation,
Screen Layout and Navigation
1-2-3
RGS-2000NG TCAS Test Set Operation Manual
139187 / Rev. 500
Page 162
Subject to Export Control. See Cover Page for details.
September 2021
3.5.8
TRANSPONDER INTERROGATION WITH CW SCREEN
The Transponder Interrogation With CW Test Mode allows the user to configure the RGS-2000NG to
transmit CW Interrogations.
SCREEN SEQUE NCE:
Transponder Screen > Transponder Test Softkey > Test Mode: Interrogation With CW
F i g u r e 1 . 2 . 3 - 1 4 5 T ra n s p o n d e r I n t e r r o ga t i o n W i t h C W S c r ee n
Screen Components
Descriptio n
Test Mode Menu
This menu selects the test Interrogation Mode.
Interrogation Type Menu
This menu selects the Interrogation Type the Single
Interrogation Mode.
PRF Field
Defines the pulse repetition rate (Pulse Repetition Frequency).
Power Field
Selects the Power.
Frequency Field
Selects the Transmitter frequency.
Settings
Interference Pulse
Not available in Interrogation with CW Mode.
Burst Tick Box
This tick box turns burst mode on or off. Burst Settings are
defined on the by pressing the Burst Settings Softkey.
Pulse Characteristic Menu
Selects the pulse to be measured.
Antenna Selection Toggle Switch
Selects the Antenna from which the pulse is measured (Top or
Bottom).
Trigger Level Slider
Selects the scope trigger level. A precise trigger level can be
defined using the Trigger Field on the Instrument Settings
Softkey Menu.
Reply Information
These fields display data contained in the intruder reply
message.